Control Technology Incubator Excellent
Temperature control, programming and documentation
For applications in food chemistry, biology and pharmaceutics that have to meet high standards of thermal safety, the Excellent class is recommended for use with the Memmert Incubator I, extensively equipped for programming and documentation. A multifunctional fuzzy controller, with permanent performance adjustments and integrated self-diagnosis system, as well as two high-quality platinum temperature sensors Pt100, ensure absolutely reliable temperature control in the incubator, as well as the highest process safety and control accuracy.
Technical specifications of Memmert Incubator Excellent
- Temperature range: 30 °C (at least 5 °C above ambient) to 70 °C
- Temperature variation (time): < ±0.05 °C at 37 °C
- Triple overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 3 °C above the setpoint. An adjustable overtemperature controller TWW protection class 3.1 (optional adjustable electronic temperature limiter TWB protection class 2) takes over control at a preset temperature in case of failure. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
- 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) with mutual sensor monitoring and operation transfer at same working temperature
- Programmable digital timer (1 min. to 999 hrs) and weekly programmer for temperature control profile with a maximum of 4 segments: Switch on with time delay, heat up, retain set temperature (dependent on setpoint, if required) and defined cooling down
- Loop (repeat function)
- Manually adjustable air flap for fresh air intake
- Speed of fan can be controlled at 10% intervals, or shut down (IF)
- Simultaneous display of all parameters, such as temperature, weekday, time, fan speed (IF) and ramp segment information
- Visual alarm if temperature is exceeded
- Function signals for stand-by, operating mode, heating and overtemperature
- Internal log memory with 1024kB to save temperature and error states, with timestamp to the minute
- Serial RS-232 communication interface (option: USB) and software Celsius to control the appliance and to read out the appliance's internal log memory
Products of interest
Peltier-Cooled Incubator
Not having a compressor creates more space, brings calm to the laboratory and provides living cultures with an extremely low-vibration environment in the IPP Peltier-cooled incubator.
CO2 incubator INCO
Ideal for research and medicine! Taking the standard INCO model and 7 additional modules, you can put together your own customised CO2 incubator.
Solutions in practice
Research for all
The incubation of bacteria in an incubator proved at an exhibition in Dublin how dangerous kisses can be!
Application Biology
Without noises and vibrations: In the laboratory of the Würzburg BEEgroup, many bees began life in a cooled Memmert Incubator IPP.