Control Technology Climatic Test Chamber
Precise temperature control in material research
Perfectly tuned control technology in the Memmert climatic test chamber CTC and in the temperature test chamber TTC guarantees real conditions for absolutely reliable climate testing and temperature testing in electronics or the automotive industries. A multifunctional fuzzy controller (P class), with permanent performance adjustments and integrated self-diagnosis system, as well as two high-quality platinum temperature sensors Pt100, ensure absolutely reliable temperature control and environment simulation, as well as the highest process safety and control accuracy.
Technical specifications
- Temperature range: -42 °C to 190 °C with a display accuracy of 0.1 °C
- High-quality and maintenance-free capacitive humidity sensor for high measurement precision and long-term stability
- Digitally controlled humidity in the range from 10 to 98% rh
- Electronically controlled refrigerant injection
- High-performance, adjustable speed air circulation with a capacity of up to 1,200 m³ per hour
- Multiple over-temperature protection: An adjustable over-temperature controller TWW protection class 3.3 takes over control in case of failure at a preset temperature. As a special safety feature we are offering an "ASF" integrated over- and under-temperature monitor, which automatically follows the setpoint value at freely selectable intervals. An alarm signal is activated if limits are crossed. In addition a TB mechanical temperature limiter switches off the heating permanently if the maximum permissible chamber temperature is exceeded
- Adjustable under-temperature alarm
- 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) communicate with each other and ensure uninterrupted error-free temperature control at constant working temperature
- Weekly timer programmable to the minute
- Simultaneous display of all temperature, humidity, weekday and time information
- Visual and acoustic alarm if temperature limits are crossed, if door is opened and if water container is empty
- Multiple-language (D/UK/E/F/I), alphanumeric text display
- Function signals for stand-by, operating mode, heating and over- /under-temperature
- Internal log memory for saving temperature and humidity as well as error states, with time stamp accurate down to the minute
- Chip card control
- USB interface for programming and storage (Ethernet optionally available at extra charge) and Celsius software to control the appliance and read out the internal log memory of the appliance
- Parallel printer interface for printout of log data via an external PCL3-compatible inkjet printer (USB possible using a converter)
Products of interest
Quality assurance in the humidity chamber
Due to the almost unlimited programming possibilities of temperature control profiles, the Memmert humidity chamber HCP provides the ideal conditions for material testing and environment simulation.
Special solutions myAtmoSAFE
For all Memmert standard products we can make pass-through appliances, customised temperature ranges, gas concentrations or air exchange, and many other myAtmoSAFE special solutions.
Solutions in practice
Application material research
Memmert constant climate chamber HPP demonstrates the behaviour of paper exposed to humidity.
Application material research
At the Fraunhofer IKTS the constant climate chamber shows that humidity has an effect on source materials and intermediate products during the manufacture of hard metal.



