Control Technology Universal Oven Basic
For precise temperature control in the laboratory
For standard temperature control tasks in science, research and industry, such as drying and heating up insensitive loads, the Basic class is recommended for use with the Universal Oven U. An electronic PID-controller with an integrated self-diagnosing system, as well as a high-quality platinum temperature sensor Pt100 ensure exact temperature control.
The double protection from overtemperature provides reliable protection for the load in the universal Memmert ovens.
Technical specifications
- Temperature range: 30 °C (at least 5°C or 10°C above ambient) to 220 °C
- Temperature variation (time): < ±0.5 °C at 150 °C
- Double overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
- A high-quality platinum temperature sensor Pt100 (DIN class A) in a 4-wire circuit (for stable long-term transmission of measurement signals)
- Programmable digital timer (1 min. to 99 hrs 59 min.)
- Manually adjustable air flap for fresh air intake
- Visual alarm if temperature is exceeded
- Function signals for stand-by, operating mode, heating and overtemperature
Products of interest
Fraunhofer TESTED DEVICE
Memmert Universal ovens UFP 500 and UFP 800 in the standard version are certified for their cleanroom suitability.
Pass-through oven for loading from both sides
Universal oven UFP 800 TS saves time when loading and reduces the danger of contamination.
Solutions in practice
Oven Drying method
Scientific examination in the laboratory demonstrates: thermally altered wood improves the tonal quality of musical instruments.
Application PCBs
The German Electrical and Electronic Manufacturers’ Association (ZVEI) recommends: Drying PCBs in a vacuum reduces drying times compared with a standard warming oven and saves energy.