控制技术 半导体恒温恒湿箱

- Memmert HPP湿度分布图(湿度范围和温度相关)
为材料测试提供精准的条件和完美的环境模拟
For a material test in a constant climate, specifically the stability test in accordance with the ICH (International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use) it is above all the precisely adjustable temperature/humidity constellations, stable over long periods, that are of crucial significance. The combination of Peltier technology for heating, cooling and dehumidifying, as applied by Memmert in its HPP constant climate chamber, is so far unique on the market.
A multifunctional fuzzy controller (P class), with permanent performance adjustments and integrated self-diagnosis system, as well as two high-quality platinum temperature sensors Pt100, ensure absolutely reliable temperature control in the constant climate chamber, as well as the highest process safety and control accuracy.
原料测试需要在一个恒定气候条件下进行,特别是符合ICH(人用药物注册技术要求国际协调会)的稳定性测试,我们的半导体恒温恒湿箱足以让所有高精度可调温湿度控制产品黯然失色,卓越的稳定性,这是最重要的,这是高所有的确确可调节温度/湿度
多功能模糊控制器(P级),可持续地进行调节并带有自诊断系统,同时有2个高品质的铂金传感器,确保在恒温恒湿箱内绝对可靠的温度控制,和最高的过程安全及控制精度。
技术参数
- 温度控制范围5°C到70°C, 显示精度和设定精度为0.1°C。
- 过温和欠温保护:可调过温/欠温保护控制器TWW(3.3级)一旦预设温度失效时将接管控制。ASF自动安全功能-独特的安全功能- 一旦超在一个设定的温度范围,温度将自动回到预设温度值,这个温度调节范围可自行设定。在过温和欠温情况下会触发声音报警,并且在过温时停止加热。另外,机械的温度限位器,TB, 在箱体超过最高允许温度10°C时将关闭加热功能
- 2个高品质4线制铂金温度传感器Pt100(长期和稳定的传输测量信号),并且传感器之间可相互沟通,确保在持续工作中不间断地无差错控温
- 数控可编程温度控制(可编程到分钟),最多可包括40个温度程序段。
- 周可编程定时时,可编程到分钟。
- 同时显示所有的参数:温度、湿度、日期、时间、程序段序号。
- 声光报警:过温/欠温、过湿/湿度不足、水箱缺水时触发。
- 多种语言显示(德语/英语/西班家/法语/意大利语)
- 功能信号:等待、操作模式、加热、冷却、加湿和除湿、过温和欠温。
- 内置1024kB存储器用于储存温度、湿度和错误信息,记录频率为每分钟。
- 芯片卡控制,包括1块32kB的存储卡(最多可存储40个程序段)
- 通过USB口,Celsius软件可以读取箱体内的数据。
- 打印并口可以连接符合PLC3标准的喷墨打印机将数据打印出来。(可转换成USB接口)
您可能感兴趣的产品
品质保证-Memmert恒温恒湿箱
由于几乎可以进行无限的温度程序控制,Memmert恒湿恒湿箱为用户的样品操作提供了一个理想的环境。
采用帕尔贴(Peltier)技术的低温培养箱 - IPP
采用帕尔贴(Peltier)技术的优点:卓越的可调节功能、极微小的温度波动、工作时的低震动以及相当的节能。