Our Peltier-cooled incubator offers unparalleled precision and control accuracy and is energy-saving and gentle. The Peltier technology minimises the risk of the sample drying out, condensation formation during the cooling down process takes place outside the working chamber, on the Peltier element.
On this page, you can find all the essential technical data on our cooled incubator with Peltier element. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE(at)memmert.com.
- forced ventilation by Peltier fan
- adaptive, fuzzy-supported microprocessor PIDcontroller, for stepless control of Peltier-heating-/ cooling system - without refrigerant
- autodiagnostic system with fault indication
- 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
- digital 7-day-programme-timer with real time clock, precise minute setting, for one set value or start of ramp operation
- integrated timer for tempering profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999 hours
- multifunctional programming via menu on 8-digit alphanumeric digital display (language to be chosen via set-up):
tempering profiles of up to 40 ramps
time- and set-point dependent operation - digital display (LED) of all set parameters, such as temperature, weekdays, time, programme status and set-up values
- resolution of display for set value and actual value 0,1 °C
- long-term logging (ring store) of all relevant data, GLP-conforming as data logger - 1024 kB
- programme stored on power failure
- parallel printer interface (incl. real-time clock with date function) for printing logging files, suitable all PCL3-compatible ink-jet printers (USB available via converter, see accessories)
- USB interface including MEMMERT Software "Celsius" for programming and documentation
- chip card control incl. one MEMoryCard XL with 32 kB (up to 40 ramps)
- incl. works calibration certificate for +10 °C and +37 °C
- with audible and visual alarm
- independently working, digitally adjustable electronic microprocessor overtemperature controller TWW protection class 3.3 maximum value for overtemperature, minimum value for undertemperature
- additional integral over- and undertemperature protection "ASF" (Auto-Safety-Function) automatically following the set value at a preset tolerance range; audible alarm is activated in case of over- or undertemperature, heating function is switched off in case of overtemperature - cooling function is switched off in case of undertemperature
- resolution of display and setting accuracy: 0,1 °C
- w x h x d: 1190 x 1620 x 840 mm
- fully insulated stainless steel doors with double locking and 4-point adjustment
- inner glass doors
- rear zinc-plated steel
- mounted on lockable castors
- w x h x d: 1040 x 1200 x 600 mm, 749 l
- fully insulated, easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing
- energy-saving Peltier heating-/cooling system (heat pump principle) integrated in the rear
- 2 stainless steel grids
- from +5 °C up to +70 °C
- temperature variation in time: < +/-0,1 °C (to DIN 12 880: 2007-05)
- temperature uniformity in chamber at 10°C and 37°C: < +/- 0.4 °C (to DIN 12 880: 2007-05)
- 230 V (+/- 10%), 50/60 Hz
- ca. 1050 W (during heating and cooling)
- net weight approx. 218 kg
- gross weight carton approx. 278 kg
- dimensions approx.:
carton w x h x d: 132 x 184 x 100 cm - the appliances must be transported upright
- 8419 8998
- Federal Republic of Germany
- DE 66812464
![]() | Perforated stainless steel shelf | E0(x) | |
![]() | Stainless Steel Grid | E3(x) | |
![]() | RS232 interface instead of USB | W6 | |
![]() | Interface Ethernet instead of USB inclusive software "Celsius Ethernet-Edition" | W4 | |
![]() | Parallel/USB converter cable with integrated power supply unit to connect PCL3-compatible HP printers with USB interface to MEMMERT units. | W1 | |
![]() | Documentation package consisting of parallel USB converter cable including PLC3-compatible HP colour inkjet printer with USB interface (HP Deskjet 6980 or successor) for direct connection of printer to Memmert unit | W2 | |
![]() | IQ check list with works test data for oven as support for validation by customer | Q1 | |
![]() | OQ check list including one free-selectable temperature distribution survey to DIN 12880: 2007-05 (size 200/300: 9 measuring points, size 400-800: 27 measuring points) with works test data for oven as support for validation by customer | Q2 | |
![]() | Software conforming to FDA "Celsius FDA-Edition" for up to 16 units | Q3 | |
![]() | Oven-linked authorisation card (User-ID-Card) - prevents undesired manipulation by unauthorised third parties | V1 | |
![]() | Computer interface RS485 (for networking a maximum of 16 ovens) instead of interface USB | V2 | |
![]() | Temperature profile write/read unit for programming via PC, for writing to and reading from the chip card, up to 40 ramps | V3 | |
![]() | Additional chip card, blank, formatted (32 kB MEMoryCard XL for a maximum of 40 ramps) | V4 |