Control Technology Incubators Basic

The basis for precise and safe temperature control

(Click on image to enlarge!)
Basic Display
(Click on image to enlarge!)
Temperature progress and monitoring devices

For standard applications with insensitive load, the Basic class is recommended. An electronic PID controller with an integrated self-diagnosis system ensures reliable and exact temperature control.

  • Temperature range: 30 °C (at least 5°C above ambient) to 70 °C
  • Temperature variation (time): < ±0.2 °C at 37 °C
  • Double overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 3 °C above the setpoint. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
  • A high-quality platinum temperature sensor Pt100 (DIN class A) in a 4-wire circuit (for stable long-term transmission of measurement signals)
  • Programmable digital timer (1 min. to 99 hrs 59 min.)
  • Manually adjustable air flap for fresh air intake
  • Visual alarm if temperature is exceeded
  • Function signals for stand-by, operating mode, heating and overtemperature