For applications that have to meet high standards of thermal safety, the Excellent class is recommended, extensively equipped for logging and documentation. A multifunctional, fuzzy-supported PID process controller, with permanent performance adjustments and integrated self-diagnosis system, ensures absolutely reliable temperature control and the highest process safety.
- Temperature range: 30 °C (at least 5 °C above ambient) to 70 °C
- Temperature variation (time): < ±0.05 °C at 37 °C
- Triple overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 3 °C above the setpoint. An adjustable overtemperature controller TWW protection class 3.1 (optional adjustable electronic temperature limiter TWB protection class 2) takes over control at a preset temperature in case of failure. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
- 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) with mutual sensor monitoring and operation transfer at same working temperature
- Programmable digital timer (1 min. to 999 hrs) and weekly programmer for temperature control profile with a maximum of 4 segments: Switch on with time delay, heat up, retain set temperature (dependent on setpoint, if required) and defined cooling down
- Loop (repeat function)
- Manually adjustable air flap for fresh air intake
- Simultaneous display of all temperature, weekday, time and ramp segment information
- Visual alarm if temperature is exceeded
- Function signals for stand-by, operating mode, heating and overtemperature
- Internal log memory with 1024kB to save temperature and error states, with timestamp to the minute
- Serial RS-232 communication interface (option: USB) and software Celsius to control the appliance and to read out the appliance's internal log memory