Control Technology Sterilisers Basic

The Basis for precise and safe temperature control

(Click on image to enlarge!)
Basic Display
(Click on image to enlarge!)
Temperature progress and monitoring devices

For sterilisation procedures without guaranteed hold time (=sterilisation time) and documentation, the Basic class is recommended. An electronic PID controller with an integrated self-diagnosis system ensures reliable and exact temperature control.
For operation with setpoint-dependent sterilisation time (guaranteed hold time) and its documentation we recommend the use of our sterilisers of the Excellent and Perfect class

  • Temperature range: 30 °C (at least 5 °C or 10 °C above ambient) to 220 °C
  • Temperature variation (time): < ±0.5 °C at 150 °C
  • Double overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
  • A high-quality platinum temperature sensor Pt100 (DIN class A) in a 4-wire circuit (for stable long-term transmission of measurement signals)
  • Programmable digital timer (1 min. to 99 hrs 59 min.)
  • Manually adjustable air flap for fresh air intake
  • Visual alarm if temperature is exceeded
  • Function signals for stand-by, operating mode, heating and overtemperature