Control Technology Sterilisers Excellent

Excellent fuzzy-supported PID control

(Click on image to enlarge!)
Excellent Display
(Click on image to enlarge!)
Temperature progress and monitoring devices

For sterilisation procedures that have to meet high standards of thermal safety, the Excellent class is recommended, extensively equipped for programming and documentation. A multifunctional, fuzzy-supported PID process controller, with permanent performance adjustments and integrated self-diagnosis system, ensures absolutely reliable temperature control and the highest process safety.  The setpoint-dependent sterilisation function Setpoint Wait absolutely guarantees precise and exactly timed sterilisation procedures (see graphic).

  • Temperature range: 30 °C (at least 5 °C or 10 °C above ambient) to 250 °C
  • Temperature variation (time): < ±0.25 °C at 150 °C
  • Triple overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. An adjustable overtemperature controller TWW protection class 3.1 (optional adjustable electronic temperature limiter TWB protection class 2) takes over control at a preset temperature in case of failure. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
  • 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) with mutual sensor monitoring and operation transfer at same working temperature
  • Programmable digital timer (1 min. to 999 hrs) and weekly programmer for temperature control profile with a maximum of 4 segments: Switch on with time delay, heat up, hold set temperature (dependent on setpoint, if required) and defined cooling down
  • Loop (repeat function)
  • Manually adjustable air flap for fresh air intake
  • Speed of fan can be adjusted at 10% intervals
  • Simultaneous display of all temperature, weekday, time, speed and ramp segment information
  • Visual alarm if temperature is exceeded
  • Function signals for stand-by, operating mode, heating and overtemperature
  • Internal log memory with 1024kB to save temperature and error states, with timestamp to the minute
  • Serial RS-232 communication interface (option: USB) and software Celsius to control the appliance and to read out the appliance's internal log memory