For standard temperature control applications with insensitive load, the Basic class is recommended. An electronic PID controller with an integrated self-diagnosis system ensures reliable and exact temperature control.
- Temperature range: 30 °C (at least 5°C or 10°C above ambient) to 220 °C
- Temperature variation (time): < ±0.5 °C at 150 °C
- Double overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
- A high-quality platinum temperature sensor Pt100 (DIN class A) in a 4-wire circuit (for stable long-term transmission of measurement signals)
- Programmable digital timer (1 min. to 99 hrs 59 min.)
- Manually adjustable air flap for fresh air intake
- Visual alarm if temperature is exceeded
- Function signals for stand-by, operating mode, heating and overtemperature