Control Technology Waterbaths Basic

(Click on image to enlarge!)
Basic display
(Click on image to enlarge!)
Temperature progress and monitoring device

The Basic controller class for standard temperature control tasks

The Basic class is recommended for standard waterbath applications. An electronic PID process control with permanent performance adjustment and integrated self-diagnosis system ensures that the set temperature in the bath is achieved and maintained exactly.

  • Temperature range: 10 °C to 95 °C plus boiling stage
  • Temperature variation (time): < ±0.5 °C at 95 °C
  • Double overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted bath temperature set at the factory is exceeded by 30°C, or if the fluid level is too low.
  • A high-quality platinum temperature sensor Pt100 (DIN class A) in a 4-wire circuit (for stable long-term transmission of measurement signals)
  • Programmable digital timer (1 min. to 99 hrs 59 min. with adjustment to the minute) for setting the Delay (delayed programme start) as well as Hold (residual operation time)
  • Visual alarm if set temperature is exceeded by more than 10 °C and if heating is switched off by temperature limiter TB