Control Technology Waterbaths Excellent

(Click on image to enlarge!)
Excellent Display
(Click on image to enlarge!)
Temperature progress and monitoring devices
(Click on image to enlarge!)
Adjustable temperature limiter and adjustable overtemperature controller

The Excellent controller class for maximum thermal safety

For safe and precise test procedures in the waterbath that also meet the strictest requirements of thermal safety and the safety of the user, the Excellent class is indispensable. A multifunctional, fuzzy-supported PID process controller with integrated self-diagnosis system ensures absolutely reliable and precise temperature control.

  • Temperature range: 10 °C to 95 °C plus boiling stage
  • Temperature variation (time): < ±0.25 °C
  • Triple overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. Electronic overtemperature monitor: With a precision of 0.1 °C, the monitoring temperature can be set up to 10 °C above the rated temperature. Depending on the type of overtemperature protection selected (can be chosen by the user in the setup menu), the adjustable temperature control TWW protection class 3.1 takes over control at a preset temperature in case of failure, or the electronic temperature limiter TWB protection class 2 switches heating off permanently if the temperature is exceeded. In addition a mechanical temperature limiter, TB, switches heating permanently off if the maximum bath temperature set at the factory is exceeded by 30 °C, or if the fluid level is too low.
  • 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) with mutual sensor monitoring and operation transfer at same working temperature
  • Possibility of calibration on the controller
  • Programmable digital timer (1 min. to 999 hrs min. with adjustment to the minute up to 99 hrs 59 min., accuracy to the hour above this ) for setting the Delay (delayed programme start) as well as Hold (residual operation time). In addition the start of the hold time can be made dependent on the set temperature being reached
  • Visual alarm and acoustic signalling of overtemperature and of too low fluid level
  • Circulation pump only for waterbath WPE 45: Circulation pump can additionally be switched on via the controller for optimal thermal uniformity in the large volume waterbath