The very precise, environment-friendly and energy-saving heating and cooling system is used in all cooled incubators of the IPP production series as well as in the Constant Climate Chambers (HPP).
Heating and cooling in one system: If a voltage is placed on a Peltier element, one side is cooled and the opposite side simultaneously heats up. Simply by reversing the polarity of the supply voltage, the hot and cold sites of the Peltier element can be swapped.
The Peltier technology works particularly economically at temperatures close to ambient temperature, since energy is only required if heating or cooling is needed, in contrast to compressor technology. Thus heating or cooling power can be fine tuned. In the same way as in a thermal pump, in heating operation thermal energy is removed from the ambient air and introduced into the chamber.
Because of the closed Peltier cooling system, no outside air is exchanged. The advantages: No de-humidification of the chamber and heavily reduced risk of contamination.