On this page, you can find all the essential technical data on the Memmert climate chamber ICH. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at sales@memmert.com.
Working temperature range | without humidity: 0 to +60 °C |
---|---|
Setting temperature range | 0 to +60 °C |
Working temperature range | from +10°C up to +60°C with humidity and/or light |
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Humidity | humidity supply with distilled water from external tank by self-priming pump |
---|---|
Humidification | humidification by hot steam generator |
Humidity | active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 % |
Light | daylight and UV-light can be programmed individually via controller |
---|---|
Light | cold white light +/- 6 %, for UV-light +/- 10 % |
Light | illumination unit consisting of 4 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm |
ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
---|---|
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable values each, temperature/humidity, for two freely selectable values CO2 |
adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, light, programme time, time zones, summertime/wintertime |
Fan | forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps |
---|
Documentation | programme stored in case of power failure |
---|---|
Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
---|---|
Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY | additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature |
Autodiagnostic system | integral fault diagnostics for temperature and humidity control |
Alarm | visual and acoustic |
Air jacket | air jacket heating system for gentle all-around heating |
---|---|
Heating concept | no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior |
Cooling | energy-saving use of CFC-free cooling/heating system (refrigerant R134a) |
Defrosting | highly efficient automatic defrosting system |
Works calibration certificate | standard values +10 °C and +37 °C, 60 % rh at +30 °C |
---|---|
Door | inner glass doors |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |
Entry port | Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back |
Internals | 2 stainless steel grid(s), electropolished |
Standard accessories | Water tank including connection hose |
Dimensions | w(A) x h(B) x d(C): 1040 x 1200 x 600 mm (d less 33 mm for fan) |
---|---|
Volume | 749 l |
Max. number of internals | 14 |
Max. loading of chamber | 200 kg |
Max. loading per internal | 30 kg |
Dimensions | w(D) x h(E) x d(F): 1224 x 1950 x 785 mm (d +56mm door handle) |
---|---|
Installation | on lockable castors |
Housing | rear zinc-plated steel |
Voltage Electrical load | 230 V, 50/60 Hz approx. 1550 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 1550 W |
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
---|---|
Ambient temperature | 15 °C to 28 °C (up to 34 °C with limited temperature range) |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | w x h x d: 1330 x 2150 x 1050 mm |
Net weight | approx. 249 kg |
Gross weight carton | approx. 319 kg |
Not all climate chambers are the same. The humidity content of the chamber load, the ambient conditions and the respective temperature-humidity working range are decisive factors in the selection of the right appliance. In the adjacent diagram, you can see the possible temperature/humidity combinations for our climate chamber ICH.
Within the respective temperature-humidity range, condensation-free permanent operation is possible. To which extent condensation may occur in the threshold range depends on the humidity content of the chamber load and the ambient conditions.
This seems not to be the appropriate device for your temperature-humidity-working range?
See the other Memmert climate chambers.