Perfectly tuned control technology guarantees ideal physiological conditions in the Memmert CO2 incubator INCO for cell cultures and tissue cultures growing in-vitro, and the active humidity control ensures short recovery times after the door has been opened. A multifunctional fuzzy controller (P class), with permanent performance adjustments and integrated self-diagnosis system, as well as two high-quality platinum temperature sensors Pt100, ensure absolutely reliable temperature control and the highest process safety and control accuracy in biotechnology, biology and medicine.
- Temperature range 20 °C (at least 8 °C above ambient) to 50 °C with a display accuracy of 0.1 °C
- Digital CO2 controller with precise infrared absorption measurement system (NDIR) with automatic zeroing
- Adjustable CO2 concentration 0.0 - 10.0 %rh
- Automatic sterilisation program via the Memmert STERICard for 4 hours at 160 °C. All sensors and fittings can remain installed during sterilisation
- Multiple overtemperature protection: An adjustable overtemperature controller TWW protection class 3.1 takes over control in case of failure at a preset temperature. ASF Automatic Safety Function - a unique safety function - for over- and undertemperature automatically following the setpoint in a freely selectable tolerance range. An alarm sound is activated in case of over- /undertemperature, and the heating is switched off in case of overtemperature. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
- Adjustable undertemperature alarm
- 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) communicate with each other and ensure uninterrupted error-free temperature control at constant working temperature
- Weekly timer programmable to the minute
- Simultaneous display of all parameters, such as temperature, CO2, humidity, weekday and time information
- Visual and acoustic alarm in case of over-/undertemperature and excess/lack of CO2, if door is opened or if gas bottle is empty
- Alphanumeric text display in different languages (D/UK/E/F/I)
- Function signals for stand-by, operating mode, heating and over/undertemperature
- Internal log memory with 1024kB to save temperature and error states, with timestamp to the minute
- Chip card control including 1 STERICard
- Serial USB-interface and software Celsius to read out the appliance's internal log memory
- Parallel printer interface for printout of log data via an external PCL3-compatible inkjet printer (USB available via converter)
Incubator
The Memmert incubator I is ideally suited for incubation with temperatures up to 70 °C - specifically for the in-vitro incubation of cell cultures, tissue cultures and bacteria cultures at 37 °C.
Sterility in Biology and biotechnology
The STERICard guarantees safe and fully automatic sterilisation (and not only disinfection) of the interior. During sterilisation, menu operation of the Memmert CO" Incubator is suspended, so that no reprogramming by accident is possible.
CO2 incubator in environmental technology
Application in-vitro incubation of bacteria cultures: Hygiene is the top priority when testing drinking water for Legionella bacteria.
Biology for all
The incubation of bacteria cultures in an incubator proved at an exhibition in Dublin how dangerous kisses can be!