If hot air sterilisation in medicine and pharmaceutics has to meet high standards of thermal safety, the Excellent class of the Memmert hot-air steriliser S is recommended, extensively equipped for programming and documentation. The setpoint-dependent sterilisation function Setpoint Wait guarantees precise and exactly timed sterilisation (see graphic). A multifunctional fuzzy controller, with permanent performance adjustments and integrated self-diagnosis system, as well as two high-quality platinum temperature sensors Pt100, ensure absolutely reliable temperature control and the highest process safety and control accuracy.
- Temperature range: 30 °C (at least 5 °C or 10 °C above ambient) to 250 °C
- Temperature variation (time): < ±0.25 °C at 150 °C
- Triple overtemperature protection: If the electronic heating control should fail, heating is switched off at a fixed level of 10 °C above the setpoint. An adjustable overtemperature controller TWW protection class 3.1 (optional adjustable electronic temperature limiter TWB protection class 2) takes over control at a preset temperature in case of failure. In addition a mechanical temperature limiter, TB, switches off heating permanently if the maximum permitted oven temperature is exceeded by approx. 10 °C.
- 2 high-grade platinum temperature sensors Pt 100 in a 4-wire circuit (for stable long-term transmission of measurement signals) with mutual sensor monitoring and operation transfer at same working temperature
- Programmable digital timer (1 min. to 999 hrs) and weekly programmer for temperature control profile with a maximum of 4 segments: Switch on with time delay, heat up, hold set temperature (dependent on setpoint, if required) and defined cooling down
- Loop (repeat function)
- Manually adjustable air flap for fresh air intake
- Speed of fan can be controlled at 10% intervals, or shut down
- Simultaneous display of all temperature, weekday, time, speed and ramp segment information
- Visual alarm if temperature is exceeded
- Function signals for stand-by, operating mode, heating and overtemperature
- Internal log memory with 1024kB to save temperature and error states, with timestamp to the minute
- Serial RS-232 communication interface (option: USB) and software Celsius to control the appliance and to read out the appliance's internal log memory
Peltier-cooled incubator
Not having a compressor creates more space, brings calm to the laboratory and provides living cultures with an extremely low-vibration environment in the IPP Peltier-cooled incubator.
IPS Storage Chamber
Our IPS temperature control storage chamber, based on heating/cooling technology using Peltier elements, is recommended specifically for long-term storage in medicine and pharmaceutics.
Research for all
The incubation of bacteria in a Memmert incubator proved at an exhibition in Dublin how dangerous kisses can be!
Beeonics in the laboratory oven
Controlled atmosphere in the laboratory: Even the beehive is 100% AtmoSAFE. An entertaining analogy.