Touch, turn & go. Quick setup in three steps. On this site we have comprehensibly gathered the corresponding equipment of the two model variants SingleDISPLAY and TwinDISPLAY.
ControlCOCKPIT with one TFT-Display
ControlCOCKPIT with two TFT-Displays
| Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, programme time, time zones, daylight-saving time | Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, programme time, relative humidity, illumination, CO2, O2, pressure, time zones, daylight-saving time |
| One temperature sensor Pt100 DIN class A in a 4-wire circuit | Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error |
| HeatBALANCE function for application specific adjustment of heating power distribution (balance) between the upper and lower heating groups in an adjustment range between -50 % and +50 % (does not apply to size 30, HPP110, IPP110plus, ICP, ICH)) | |
| Software AtmoCONTROL for data read out and programming of the data logger via Ethernet interface (temporary trial version as free download). USB pen drive with software AtmoCONTROL available as accessory | Software AtmoCONTROL for programming, managing and transferring of profiles via Ethernet interface or USB port. |
| ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function | |
| Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week) | |
| Ethernet interface on the rear of the appliance for reading out the protocol log and for online logging | Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading programmes and for online logging |
| Double overtemperature protection: Electronic temperature monitoring with freely adjustable monitoring temperature, for models U, I, S with option A6 TWW/TWB (protection class 3.1 or 2), mechanical temperature limiter TB acc. to DIN 12 880 | Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2 resp. 3.3 for units with active cooling) and mechanical temperature limiter TB (protection class 1) acc. to. DIN 12 880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature and also for all other parameters such as relative humidity, CO2 |
| Structured stainless steel housing, rear of zinc-plated steel, ControlCOCKPIT for operation and adjustment of all parameters | |
| High-temperature connectors on the rear of the appliance for single-phase power connection according to country specific systems and IEC standards | |
| Internal data logger with a storage capacity of at least 10 years | |
| German, English, French, Spanish language settings available on the ControlCOCKPIT | |
| Digital backwards counter with target time setting, adjustable between 1 minute and 99 days, 23 hours | |
| The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points - optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber | |
| Adjustment of three calibration values for temperature and additional appliance specific parameters directly at the ControlCOCKPIT (e. g. relative humidity) | |