The test was performed once without any load, once with mixed load and twice at full load (metallic instruments partially wrapped in plastic or aluminium foil as well as glass containers).
In the SNE 200 steriliser (with a chamber volume of 32 litres), a total of 16 measuring sensors monitored the temperature.
In the SFP 500 steriliser, the larger inner chamber volume of 108 litres made it possible to attach 27 measuring sensors at different positions in the chamber and on the chamber load.
During each measurement, the minimum temperature of 180 °C was reached and maintained for at least 30 minutes at each measuring point.
With this validation report, the unique competitive edge that the setpoint wait function of the Memmert hot-air steriliser offers, has finally been validated by an independent laboratory.
In addition to the standard Pt100 temperature sensor, up to three external temperature sensors can be integrated in the Memmert hot-air steriliser programme sequence and thus ensure that the sterilisation time does not start before the set temperature has been reached at all measurement points.