Download here the dimensioned model sketch and extensive technical data. For long-term storage of samples, the Memmert storage chamber is recommended. Flexibility and technical features of our appliances meet all possible needs. Put us to the test!
Working temperature range/ Setting temperature range | +14 to +45 °C |
---|---|
Setting accuracy temperature | 0.1 °C |
Temperature sensor | 1 Pt100 sensor DIN class A in 4-wire-circuit |
Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
---|---|
ControlCOCKPIT | SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display |
Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT | the process time does not start until the set temperature is reached |
Calibration | three freely selectable temperature values |
adjustable parameters | temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime |
Interface | Ethernet LAN |
---|---|
Documentation | programme stored in case of power failure |
Programming | AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand). |
Temperature control | adjustable electronic overtemperature monitor and mechanical temperature limiter |
---|---|
Autodiagnostic system | for fault analysis |
Peltier | energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle) |
---|
Internals | 2 stainless steel grid(s), electropolished |
---|---|
Works calibration certificate | incl. works calibration certificate for +18°C and +25°C |
Door | inner glass doors |
Door | fully insulated stainless steel doors with 2-point locking (compression door lock) |
Dimensions | w(A) x h(B) x d(C): 1040 x 1200 x 600 mm (d less 10 mm for fan - Peltier) |
---|---|
Volume | 749 l |
Max. number of internals | 14 |
Max. loading of chamber | 200 kg |
Max. loading per internal | 30 kg |
Dimensions | w(D) x h(E) x d(F): 1224 x 1720 x 856 mm (d +56mm door handle) |
---|---|
Installation | on lockable castors |
Housing | rear zinc-plated steel |
Voltage Electrical load | 230 V, 50/60 Hz approx. 550 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 550 W |
Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
---|---|
Ambient temperature | 16 ºC to 28 ºC |
Humidity rh | max. 70 %, non-condensing |
Altitude of installation | max. 2,000 m above sea level |
Overvoltage category | II |
Pollution degree | 2 |
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | w x h x d: 1330 x 1910 x 1050 mm |
Net weight | approx. 230 kg |
Gross weight carton | approx. 301 kg |