IVF module for ICO105med

Ideally suited for in vitro fertilisation. Controlled CO2 and O2 concentration as well as gentle cultivation in Petri dishes in separate slide-in units.

ICO105med IVF
Production discontinued:
31.12.2020

IVF module for ICO105med

Standard delivery IVF module for ICOmed

On this page, you can find all the essential technical data on the Memmert CO2 incubator ICOmed with IVF module. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at sales@memmert.com.

Temperature

Working-temperature range 5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting temperature range +18 to +50 °C
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature variation in chamber at + 37 °C +/- 0.3 K
Temperature fluctuations with time at 37 °C +/- 0.1 K

Humidity

Humidity control (standard) Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
Setting accuracy humidity 0.5 % rh
Setting range active humidity control (with option K7) 40 to 97 % rh and rh-Off

Control of standard components

CO2 control Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO2 0 to 20 % CO2
Setting accuracy CO2 0,1%
Variation in time CO2 +/- 0.2 % CO2
Adjustment range O2 1 to 20 % O2
Setting accuracy O2 0.1 % O2

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
Function SetpointWAIT the process time does not start until the set temperature is reached
Adjustable parameters temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime

Communication

Interface Ethernet LAN, USB
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and CO2
Alarm visual and acoustic

Heating concept

6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid condensation

Standard equipment

Standard accessories Membrane filter (in order to remove impurities and pollutants, all incoming gases pass through a membrane filter before they reach the chamber)
IVF-module patented, consisting of 8 slide-in units, a total of 16 special racks with indentations for 16 Petri dishes (60 mm diam.) resp. 32 Petri dishes (35 mm diam.), 2 racks with indentations for 3 special media tubes each; racks with indentations for 4-well dishes on demand;
Humidity control (standart) Active microprocessor control for humidifying and dehumidifying (40 – 97 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times while avoiding condensate formation. Humidity supply with distilled water (from an external tank) by a self-priming pump; integral bacteria block by generating hotsteam, dehumidifying via sterile filter
Door fully insulated stainless steel door with 2-point locking (compression door lock)
Door inner glass door with opening (8 mm Ø) to take gas sample
works calibration certificate works calibration certificate (measuring point chamber centre) at +37 °C, 5 %, 6 % and 7 % CO2 as well as 90 % rh; 5 % O2 for IVF unit equipped with option T6

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 35 mm for fan)
Interior material 1.4301 (ASTM 304), corrosion resistant
Volume 107 l
Max. number of internals 6
Max. loading of chamber 90 kg
Max. loading per internal 15 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 719 x 850 x 591 mm (d +56mm door handle)
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1300 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1300 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 10 °C to 35 °C
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 800 x 1030 x 800 mm
Net weight approx. 75 kg
Gross weight carton approx. 100 kg

Standard units are safety-approved and bear the test marks

CE LGA
UL
VDE

More product series of Archive

Cooled storage incubator

Cooled storage incubator

Sample breeding and storage in microbiology, food, cosmetics or pharmaceutics: The cooled storage incubator with Peltier elements lowers operating costs as well as CO2 emissions.

Oilbath

Oilbath

Absolute safety for laboratory and samples, even at high temperatures through corrosion-resistant stainless steel, precise electronics and multiple temperature protection.

Constant climate chamber

Constant climate chamber

Our constant climate chamber is tailored to environmental simulation, material testing and stability testing in accordance with the ICH guidelines.

Cleanroom constant climate chamber

Cleanroom constant climate chamber

Fraunhofer TESTED DEVICE in the standard version. The cleanroom-compliant constant climate chamber is ideal for environmental simulation, climate testing and stability testing.

Peltier-cooled incubator

Peltier-cooled incubator

Microbiology, zoology, food, cosmetics or pharma industry: the energy-saving cooled incubator with Peltier elements heats up and cools down seamlessly in one system.