Archive

Peltier-cooled incubator IPP55

Microbiology, zoology, food, cosmetics or pharma industry: the energy-saving cooled incubator with Peltier elements heats up and cools down seamlessly in one system.

IPP55
Production discontinued:
31.12.2020

Peltier-cooled incubator IPP55

Standard delivery Peltier-cooled incubator

With the help of our model selection, with dimensioned model sketches and extensive technical data for download, you will find your perfect Peltier-cooled incubator. For large volumes in conjunction with rapid temperature changes, the Memmert compressor-cooled incubator is recommended. Flexibility and technical features of our appliances meet all possible needs. Put us to the test!

Temperature

Setting temperature range 0 to +70 °C
Working temperature range without light: from 0 (at least 20 below ambient temperature) to +70°C
Setting accuracy temperature 0.1 °C
Temperature sensor 1 Pt100 sensor DIN class A in 4-wire-circuit

Control technology

Language setting German, English, Spanish, French, Polish, Czech, Hungarian
ControlCOCKPIT SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime

Ventilation

Convection forced ventilation by Peltier fan

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).

Safety

Temperature control adjustable electronic overtemperature monitor and mechanical temperature limiter
Autodiagnostic system for fault analysis

Heating concept

Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)

Standard equipment

Door fully insulated stainless steel door with 2-point locking (compression door lock)
Internals 1 stainless steel grid(s), electropolished
Works calibration certificate for +10°C and +37°C
Door inner glass door

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 400 x 400 x 330 mm (d less 10 mm for fan - Peltier)
Volume 53 l
Max. number of internals 4
Max. loading of chamber 80 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 585 x 784 x 586 mm (d +56mm door handle)
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 275 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 275 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 730 x 950 x 670 mm
Net weight approx. 52 kg
Gross weight carton approx. 71 kg

Standard units are safety-approved and bear the test marks

CE
EAC Eurasian Conformity

More product series of Archive

IVF module for ICOmed

IVF module for ICOmed

Ideally suited for in vitro fertilisation. Controlled CO2 and O2 concentration as well as gentle cultivation in Petri dishes in separate slide-in units.

Waterbath

Waterbath

The combination of corrosion-resistant stainless steel, precise electronics and multiple temperature protection guarantees the highest level of safety in the laboratory.

Cooled storage incubator

Cooled storage incubator

Sample breeding and storage in microbiology, food, cosmetics or pharmaceutics: The cooled storage incubator with Peltier elements lowers operating costs as well as CO2 emissions.

Oilbath

Oilbath

Absolute safety for laboratory and samples, even at high temperatures through corrosion-resistant stainless steel, precise electronics and multiple temperature protection.

Constant climate chamber

Constant climate chamber

Our constant climate chamber is tailored to environmental simulation, material testing and stability testing in accordance with the ICH guidelines.

More product categories

Heating / drying ovens

Incubators

Climate chambers

Waterbaths