Cleanroom constant climate chamber HPP1060

Fraunhofer TESTED DEVICE in the standard version. The cleanroom-compliant constant climate chamber for environmental simulation, climate testing and stability testing.

HPP1060
New

Cleanroom constant climate chamber HPP1060

Standard delivery Cleanroom constant climate chamber

On this page, you can find all the essential technical data on the Memmert stability chamber HPP. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at sales@memmert.com.

Temperature

Setting temperature range without light, with humidity: +5 to +70 °C
Setting temperature range without light, without humidity: 0 to +70 °C
Working temperature range without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C
Working temperature range without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
Working temperature range without light: from 0 (at least 20 below ambient temperature) to +70°C
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity

Setting range humidity 10 - 90 % rh
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidification humidification by hot steam generator
Dehumidification dehumidification by cold trap using the Peltier technology
Setting accuracy humidity 0.5 % rh

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable values each, temperature and humidity

Communication

Interface Ethernet LAN, USB
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic

Heating concept

Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)

Standard equipment

Works calibration certificate for +25 °C/40 % rh and +40 °C/75 % rh
Internals 2 stainless steel grid(s), electropolished
Standard accessories Water tank including connection hose (110-750: 2,5 liter, 1060-2200: 10 liter)
Door fully insulated stainless steel doors with 2-point locking (compression door lock)

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 1040 x 1200 x 850 mm (d less 10 mm for fan - Peltier)
Volume 1060 l
Max. number of internals 14
Max. loading of chamber 200 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 1224 x 1720 x 1107 mm (d +56mm door handle)
Installation on lockable castors
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1600 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1600 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 1370 x 1970 x 1300 mm
Net weight approx. 260 kg
Gross weight carton approx. 424 kg

Standard units are safety-approved and bear the test marks

CE

More product series of Climate chambers

Constant climate chamber

Constant climate chamber

Maximum energy efficiency in continuous operation - tailor-made for stability studies according to ICH guidelines, stability tests for cosmetics and foods as well as environmental testing and material testing

Climate chamber

Climate chamber

Unparalleled homogeneity of temperature and humidity! For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP – GCCP.

Humidity chamber

Humidity chamber

Active humidity control creates a controlled environment for accelerated life tests and 85/85 tests.

Environmental test chambers

Environmental test chambers

Memmert environmental test chambers guarantee the perfect atmosphere for climate and temperature tests

More product categories

Heating / drying ovens

Incubators

Waterbaths

Archive