Compressor-cooled incubator ICP450

Do you require different ramps for incubating, breeding or storing? Do you need a larger chamber volume and at the same time quick temperature changes?

ICP450

Compressor-cooled incubator ICP450

Standard delivery Compressor-cooled incubator

With the help of our model selection, with dimensioned model sketches and extensive technical data for download, you can find the right cooled incubator for your needs. For small volumes and for work predominantly in the proximity of the ambient temperature, the Memmert Peltier-cooled incubator is recommended. Flexibility and technical features of our appliances meet all possible needs. Put us to the test!

Properties

Temperature

Working-temperature range from -12°C up to +60°C (Optimum performance of cooling aggregate at an ambient temperature of +16°C up to +32°C. Not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over.)
Setting accuracy temperature 0.1°C
Temperature 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), fan speed, programme time, time zones, summertime/wintertime

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system for fault analysis
Alarm visual and acoustic

Heating concept

Air jacket air jacket heating system for gentle all-around heating
Heating concept no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Cooling energy-saving use of CFC-free cooling/heating system (refrigerant R134a)
Defrosting highly efficient automatic defrosting system

Standard equipment

Internals 2 stainless steel grid(s), electropolished
Scope of delivery incl. works calibration certificate for +10°C and +37°C
Door inner glass doors
Door fully insulated stainless steel doors with 2-point locking (compression door lock)

Stainless steel interior

Volume 449 l
Dimensions W x H x D in mm w(A) x h(B) x d(C): 1040 x 720 x 600 mm
Max. number of internals 8
Max. loading of chamber 200 kg
Max. loading per internal 30 kg

Textured stainless steel casing

Installation on lockable castors
Dimensions w(D) x h(E) x d(F): 1224 x 1613 x 784 mm
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1200 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1200 W

Packing/shipping data/Setting Up

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton B x H x T: 1330 x 1700 x 1050 mm
Net weight approx. 217 kg
Gross weight carton approx. 282 kg

More product series of Incubators

Incubator Im

The incubator Im is a Class I medical device in accordance with the EU directive 93/42/EEC.

CO<sub>2</sub> Incubator

CO2 Incubator

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Cooled storage incubator

Cooled storage incubator

Sample breeding and storage in microbiology, food, cosmetics or pharmaceutics: The cooled storage incubator with Peltier elements lowers operating costs as well as CO2 emissions.

CO<sub>2</sub> Incubator (med)

CO2 Incubator (med)

Protection for in vitro growing cell cultures, tissue cultures and bacteria cultures in a controlled and physiologically ideal environment.

Peltier-cooled incubator

Peltier-cooled incubator

Microbiology, zoology, food, cosmetics or pharma industry: the energy-saving cooled incubator with Peltier elements heats up and cools down seamlessly in one system.

Incubator

Incubator

The incubator I is at home everywhere in the world of research, medicine, pharmaceutics and food analytics, as well as food chemistry.

More product categories

Heating / drying ovens

Climate chambers

Water and oilbaths