Environmental test chambers TTC256

Memmert environmental test chambers guarantee the perfect atmosphere for climate and temperature tests

TTC256

Environmental test chambers TTC256

Standard delivery Environmental test chambers

On this page, you can find all the essential technical data on the Memmert environmental chamber CTC/TTC for climate testing and temperature testing. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.

Properties

Temperature

Temperature range without humidity from -42°C up to +190°C
Temperature distribution (spatial) +/- 0.5 up to 2K
Temperature 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Display resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED)

Humidity

Dehumidification 2 x 10 l tanks as condensate collector

Control technology

Controller Electronic microprocessor temperature controller with auto-diagnostic system
Controller digital display of all set parameters, such as temperature, weekdays, time, fan speed, air flap position, programme status and set-up values - language to be chosen via set-up
Timer integrated timer for tempering profiles of up to 40 ramps each, each segment adjustable from 1 min. to 999 hrs.
Calibration three freely selectable temperature values

Ventilation

Fan high-performance air fan in working chamber
Fan control depending on operation status automatic adaption of fan speed resp. manual adjustment from 10 to 100 %

Communication

Interface USB USB-interface incl. Memmert software "Celsius" for programming and documentation
Interface Printer parallel printer interface (incl. real time clock with date function) for all PCL3-compatible ink jet printers for GLP-conforming documentation
Documentation integrated ring memory as data logger for GLP-conforming long-term documentation of all relevant parameters - 1024 kB
Documentation programme stored in case of power failure
Programming chip-card control incl. 1 MEMoryCard XL with 32 kB storage capacity (max. 40 ramps)

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, adjustable at controller
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature control
Alarm with visual and acoustic alarm in case of over-/under temperature and open door

Heating concept

Ring heater high-performance ring heaters with optimised air circulation
Door door heating to avoid condensation
Cooling twin compressor; chlorine-free refrigerant R404A

Standard equipment

Works calibration certificate standard values +30°C u. 60 % rh
Door fully insulated stainless steel door with double-locking and 4-point adjustment, heated
Interior easy-to-clean stainless steel interior, materil 1.4301 (ASTM 304), hermetically welded
Internals 1 stainless steel grid

Stainless steel interior

Volume 256 l
Dimensions W x H x D in mm w(A) x h(B) x d(C): 640 x 670 x 597 mm
Max. number of grids/shelves 6
Max. loading per grid/shelf 25 kg

Textured stainless steel casing

Installation on lockable castors
Dimensions w(D) x h(E) x d(F): 898 x 1730 x 1100 mm
Housing rear zinc-plated steel

Electrical data

Electrical load 400 V, 3 phases, 50/60 Hz / approx. 7000 W

Packing/shipping data

the appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton B x H x T: 1020 x 1910 x 1310 mm
Net weight approx. 337 kg
Gross weight carton approx. 463 kg

Temperature-humidity working range CTC

Not all climate chambers are the same. The humidity content of the chamber load, the ambient conditions and the respective temperature-humidity working range are decisive factors in the selection of the right appliance. In the adjacent diagram, you can see the possible temperature/humidity combinations for our climatic test chamber CTC.



This seems not to be the appropriate device for your temperature-humidity-working range?
See the other Memmert climate chambers.

More product series of Climate chambers

Humidity chamber

Humidity chamber

High-precision control technology creates controlled and physiologically ideal surroundings for the perfect environment simulation in building physics, electronics, biology, zoology and botany.

Climate chamber

Climate chamber

Unparalleled homogeneity of temperature and humidity! For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP – GCCP.

Climate chamber for keeping mice

Climate chamber for keeping mice

„Specific Pathogen Free” (SPF) keeping of mice and rats in compliance with GV Solas guidelines.

Constant climate chamber

Constant climate chamber

Our constant climate chamber is tailored to environmental simulation, material testing and stability testing in accordance with the ICH guidelines.

More product categories

Heating / drying ovens

Incubators

Water and oilbaths