Oilbath ONE 14

Absolute safety for laboratory and samples, even at high temperatures through corrosion-resistant stainless steel, precise electronics and multiple temperature protection.

ONE 14

Oilbath ONE 14

Standard delivery Oilbath

On this page, you can find all the essential technical data on the Memmert oil bath. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.

Properties

Temperature

Temperature range in °C min. 5°C above ambient up to +200°C
resolution of display and setting accuracy +0,1°C below 99,9°C, 1°C above 100°C

Control of standard components

Controller digital display (LED) of all set parameters, such as temperature and alarm values (0,1°C resolution) as well as time values
Timer digital timer from 1 min. up to 999 hours for: ON (continuous operation), DELAYED ON, HOLD or HOLD set-temperature dependent with guaranteed holding time
Calibration on controller

Safety

Temperature control 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Temperature control mechanical temperature limiter TB protection class 1 switching the heating off at approx. 30°C above max. temperature of the bath
Temperature control independently working, electronic overtemperature temperature limiter TWB protection class 2
Temperature control in case of overtemperature due to failure, the heating is switched off at approx. 10°C above the set temperature (fixed value)
Autodiagnostic system fuzzy-supported PID microprocessor controller with integrated autodiagnostic system with fault indication
Alarm visual and acoustic alarm at programme end and as input acknowledgement as well as in case of low liquid level, heating is switched off automatically

Heating concept

Heating Baths corrosion-proof large-area heating on three sides

Stainless steel interior

Interior easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing, material no. 1.4301 (ASTM 304), laser-welded
Volume 14 l
Dimensions W x H x D in mm w(A) x h(B) x d(C): 350 x 290 x 140 mm
Liquid level min. (H) 97 mm
Liquid level max. (I) 120 mm

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 578 x 436 x 238 mm

Electrical data

Voltage 230 V, 50/60 Hz
Electrical load approx. 1800 W (during heating)

Packing/shipping data/Setting Up

the appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton B x H x T: 670 x 530 x 400 mm
Net weight approx. 15 kg
Gross weight carton approx. 21 kg