Constant climate chamber HPP1060

Our constant climate chamber is tailored to environmental simulation, material testing and stability testing in accordance with the ICH guidelines.

HPP1060
New

Constant climate chamber HPP1060

Standard delivery Constant climate chamber

On this page, you can find all the essential technical data on the Memmert stability chamber HPP. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.

Temperature

Working temperature range from 0°C to +70°C
Setting accuracy temperature 0.1 °C
Temperature 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity

Setting range humidity 10 - 90 % rh
Setting accuracy humidity 0.5 % rh
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidification humidification by hot steam generator
Dehumidification dehumidification by cold trap using the Peltier technology

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable values each, temperature and humidity

Communication

Interface Ethernet LAN, USB
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic

Heating concept

Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)

Standard equipment

Works calibration certificate for +25 °C/40 % rh and +40 °C/75 % rh
Internals 2 stainless steel grid(s), electropolished
Door Stainless steel doors with glass sectors, fully heated inner glass panes integrated in the full-sight glass door with 2-point locking (compression door lock)
Standard accessories Water tank including connection hose

Stainless steel interior

Volume 1060 l
Dimensions w(A) x h(B) x d(C): 1040 x 1200 x 850 mm
Max. number of internals 14
Max. loading of chamber 200 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Installation on lockable castors
Dimensions w(D) x h(E) x d(F): 1224 x 1661 x 1139 mm
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1500 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1500 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 1370 x 1970 x 1300 mm
Net weight approx. 260 kg
Gross weight carton approx. 424 kg

Standard units are safety-approved and bear the test marks

CE

Temperature-humidity working range HPP

Not all climate chambers are the same. The humidity content of the chamber load, the ambient conditions and the respective temperature-humidity working range are decisive factors in the selection of the right appliance. In the adjacent diagram, you can see the possible temperature/humidity combinations for our constant climate chamber HPP.

Within the respective temperature-humidity range, condensation-free permanent operation is possible. To which extent condensation may occur in the threshold range depends on the humidity content of the chamber load and the ambient conditions.

This seems not to be the appropriate device for your temperature-humidity-working range?
See the other Memmert climate chambers.

More product series of Climate chambers

Humidity chamber

Humidity chamber

High-precision control technology creates controlled and physiologically ideal surroundings for the perfect environment simulation in building physics, electronics, biology, zoology and botany.

Climate chamber

Climate chamber

Unparalleled homogeneity of temperature and humidity! For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP – GCCP.

Constant climate chamber for keeping mice

Constant climate chamber for keeping mice

Safe and stable air humidity in compliance with GV Solas guidelines.

Environmental test chambers

Environmental test chambers

Memmert environmental test chambers guarantee the perfect atmosphere for climate and temperature tests