IN55plus

The incubator I is perfect for the world of research, medicine, pharmaceutics and food analytics, as well as food chemistry.

IN55plus

IN55plus

Standard delivery Incubator

The heating of this incubator is optimally tuned for natural convection and valuable chamber loads for research, pharmaceutics, medicine and food chemistry are warmed up very carefully.

On this page, you can find all the essential technical data on our incubator. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at  sales@memmert.com.

Temperature

Setting temperature range +20 to +80 °C
Working temperature range at least 5 above ambient temperature to +80 °C
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Sterilisation fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load

Ventilation

Convection natural convection
Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
Vent vent connection with restrictor flap

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Temperature control overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system for fault analysis
Alarm visual and acoustic

Standard equipment

Works calibration certificate incl. works calibration certificate for +37°C
Door fully insulated stainless steel door with 2-point locking (compression door lock)
Door inner glass door
Internals 1 stainless steel grid(s), electropolished

Stainless steel interior

Interior easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume 53 l
Dimensions w(A) x h(B) x d(C): 400 x 400 x 330 mm
Max. number of internals 4
Max. loading of chamber 80 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 585 x 784 x 514 mm (d +56mm door handle)
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1000 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 900 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installation max. 2,000 m above sea level
Ambient temperature +5 °C to +40 °C
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 730 x 950 x 670 mm
Net weight approx. 57 kg
Gross weight carton approx. 76 kg

Standard units are safety-approved and bear the test marks

CE
VDEGS
UL
EAC Eurasian Conformity

More product series of Incubators

Peltier-cooled incubator

Peltier-cooled incubator

Sets eco standards for cultivation below room temperature: unmatched energy efficiency, best values for heat-up, cool-down and recovery times.

CO<sub>2</sub> incubator

CO2 incubator

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Incubator Im

Incubator Im

The incubator Im is a Class I medical device.

CO<sub>2</sub> incubator ICOmed

CO2 incubator ICOmed

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Compressor-cooled incubator

Compressor-cooled incubator

Ideal for ramp operation with rapid temperature changes during incubation, breeding or storage.

More product categories

Heating / drying ovens

Climate chambers

Waterbaths

Archive

m360