Universal Oven Um UN450m

The heating oven Um is a Class I medical device in accordance with the EU directive 93/42/EEC.

UN450m
Classified medical device

Universal Oven Um UN450m

Standard delivery Universal Oven Um

This universal oven made of high-quality hygienic, easy to clean stainless steel leaves nothing to be desired in terms of ventilation technology, control technology, overtemperature protection and perfectly tuned heating technology.

Control technology

Language setting German, English, Spanish, French, Polish, Czech, Hungarian
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values

Ventilation

Vent vent connection with restrictor flap

Communication

Documentation programme stored in case of power failure

Safety

Autodiagnostic system for fault analysis

Standard equipment

Internals 2 stainless steel grid(s), electropolished
Works calibration certificate Calibration at +160°C
Door fully insulated stainless steel doors with 2-point locking (compression door lock)

Stainless steel interior

Interior easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume 449 l
Dimensions w(A) x h(B) x d(C): 1040 x 720 x 600 mm
Max. number of internals 8
Max. loading of chamber 300 kg
Max. loading per internal 30 kg

Textured stainless steel casing

Installation on lockable castors
Dimensions w(D) x h(E) x d(F): 1224 x 1247 x 784 mm
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 400 V, 3 phases, 50 Hz approx. 5800 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installation max. 2,000 m above sea level
Ambient temperature +5 °C to +40 °C
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 1330 x 1440 x 1050 cm
Net weight approx. 161 kg
Gross weight carton approx. 227 kg

Standard units are safety-approved and bear the test marks

VDE
GS
CE
EAC Eurasian Conformity
UL

More product series of Heating / drying ovens

Universal Oven

Universal Oven

Precise drying, heating, ageing, burn-in and hardening in research, science, industry and quality assurance.

Steriliser

Steriliser

Safer than disinfection, hot-air sterilisation of containers and instruments

Pass-through oven

Pass-through oven

The Memmert pass-through oven saves time when loading and reduces the danger of contamination, especially when directly transporting the chamber load between the grey room and the clean room.

Paraffin oven

Paraffin oven

For many years, and with great precision, our paraffin oven UNpa has served users in science and research in sample preparation.

Cleanroom drying oven

Cleanroom drying oven

The cleanroom-compliant drying oven for microelectronics, medical technology, pharmaceutics and food technology.

Vacuum oven

Vacuum oven

Drying food, cosmetics, clocks, books, PCBs or injection moulds: Design your own VO vaccuum oven according to your wishes!

Cooled vacuum oven

Cooled vacuum oven

For gentle drying of bacteria and starter cultures or simulation of intercontinental flights.

Universal oven dishwasher

Universal oven dishwasher

Air circulation heating oven in accordance with EN 50242/EN 60436: 2013-08

More product categories

Incubators

Climate chambers

Water and oilbaths