Options / Accessories

for CO2 Incubator ICO50med

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Option

Option Order No.
Voltage 115 V, 50/60 Hz (no extra cost) X2
Battery-buffered ControlCOCKPIT. Uninterrupted supply for the entire display unit (ControlCOCKPIT) and therefore complete documentation of all parameteres even when there is a power failure. The CO2 parameter is continously regulated. C2
Two gas connections with quick release connectors, automatic switch-over of gas cylinders; incl. two CO2 connection sets: hose with coupling and clamp T1
Electropolished interior T2
Active microprocessor control for humidifying and dehumidifying (40 - 97 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times while avoiding condensate formation. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter K7
Control of oxygen concentration by N2 inlet; adjustment range 1 % up to 20 % O2; setting accuracy 0.1 % (requires option K7) T6
Peltier cooling unit enables a working temperature of 37 °C even at higher ambient temperatures of up to 35 °C K5
Capacitive humidity sensor for measuring and displaying the relative humidity K6
Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone stopper, at the back, centre right; not available for ICO50med with active humidity control (option K7) or humidity display (option K6) F7
4 – 20 mA current loop interface O2 controller, actual value (0 – 25 % O2 4 – 20 mA) V1
4 – 20 mA current loop interface CO2 controller, actual value (0 – 25 % CO2 4 – 20 mA) V9
4 – 20 mA current loop interface Humidity controller, actual value (0 – 100 % rh 4 – 20 mA) V7
4 – 20 mA current loop interface Temperature controller, actual value (0 to +70 °C 4 – 20 mA) V3
Works calibration certificate for 5 %, 7 % and 10 % CO2 (measured at +37 °C) special works calibration certificates upon request D00106
Works calibration certificate for one (freely selectable) temperature, humidity, CO2 and O2 value according to customer specification (requires T6) D00143
Works calibration certificate for one (freely selectable) temperature, humidity and CO2 value according to customer specification (requires K7) D00131
Start-up of ICOmed incubators and brief training (D, A, CH only), through Memmert service, not subject to discount K9
Door hinged on the left B8
Potential-free contact (24V/2A) with socket 3 pin,
according to NAMUR NE 28 for external monitoring
(setpoint of temperature and CO<sub>2</sub> is reached) Potential-free contact (24V/2A) with socket 3 pin, according to NAMUR NE 28 for external monitoring (setpoint of temperature and CO2 is reached) H5
Potential-free contact for combination error message (e.g. supply failure, sensor fault, fuse) Potential-free contact for combination error message (e.g. supply failure, sensor fault, fuse) H6
MobileALERT for up to 4 alarm notifications; standard: temperature and CO2 alarm, additionally humidity alarm (when equipped with option K7) and O2 alarm (when equipped with option T6) C4
MobileALERT, notification by SMS in case of any error or alarm of the device. Requires option H6 “floating contact for alarm” C3

Accessory

Accessory Order No.
Additional perforated stainless steel shelf E35160
Additional water dish (not applicable with option K7) B38737
Central water supply with filter cartridges for connection to the domestic water supply, only in combination with option K7. Product information on demand ZWVR6
Central water supply, without filter cartridges for connection to the domestic water supply (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7) only in combination with option K7. Product information on demand ZWVR7
HEPA-filter for chamber according to EN 1822, packed in sterile condition, incl. fixing unit B38739
Guarantee extension by 1 year GA3Q5
IVF module for ICO50med: patented, consisting of 6 slide-in units, a total of 12 special racks with indentations for 12 Petri dishes (60 mm diam.) resp. 24 Petri dishes (35 mm diam.), 2 racks with indentations for 3 special media tubes each; racks with indentations for 4-well dishes on demand; only for ICO50med with the options K7 and F7; works calibration certificate (measuring point chamber centre) at +37 °C, 5 %, 6 % and 7 % CO2 as well as 90 % rh; 5 % O2 for IVF unit equipped with option T6 B44128
Holder for test tube (only in combination with IVF-module) E37069
Holder for Petri dishes square (only in combination with IVF-module) E37308
Holder for Petri dishes round (only in combination with IVF-module) E37026
Magnetic foil, can be labelled with a non-permanent board marker (only in combination with IVF-module) E07620
USB-Ethernet adapter E06192
Ethernet connection cable 5 m for computer interface E06189
USB User-ID stick (with User-ID licence): Oven-linked authorisation licence (User-ID-programme) on Memory-stick, prevents undesired manipulation by unauthorised third parties. When reordering please specify serial number (only for units with TwinDISPLAY) B33170
Set of height adjustable feet (4 pcs). Standard on models ICOmed B29768
Stacking set (4 pcs) for stacking of appliances of same size (not for models 160, 260, 450, 750, 1060, ICH110, ICH110L and ICH110C) Stacking set (4 pcs) for stacking of appliances of same size (not for models 160, 260, 450, 750, 1060, ICH110, ICH110L and ICH110C) B29744
Subframe (130 mm high); sizes 150/240: only in combination with the corresponding stacking sets for stacked appliances. B33507
Subframe (622 mm high) adjustable in height (sizes 150/240: should not be used for 2 stacked units) B33504
CO2 pressure reducing valve DIN 8546, incl. Gas cylinder monitor E02087
N2 pressure reducing valve to DIN EN ISO 2503, incl. gas cylinder monitor (requires option T6) E06162
Integration of one additional unit (up to max. 15 units) into an already existent FDA-software licence FDAQ2
IQ/OQ document with device-specific works test data for one freeselectable CO2 and temperature value, incl. temperature distribution survey at Memmert for 27 measuring points to DIN 12880:2007-05, PQ check list as support for validation by customer (models ICOmed). D38898
IQ/OQ document with device-specific works test data for one free-selectable CO2, humidity and temperature value, incl. temperature distribution survey at Memmert for 27 measuring points to DIN 12880:2007-05, PQ check list as support for validation by customer (models ICH C and ICOmed, on models ICOmed a free-selectable humidity value is only possible with option K7). D38897
Software conforming to FDA “AtmoCONTROL“. Meets the requirements for the use of electronically stored data sets and electronic signatures as laid down in Regulation 21 CFR Part 11 of the US Food and Drug Administration (FDA). Base licence for the control of one unit (only for units with TwinDISPLAY). Respective IQ/OQ documents available in German and English language (without surcharge). FDAQ1
IQ document with device-specific works test data, OQ/PQ check list as support for validation by customer D00124
External measuring instrument with sensors with additional measuring head for temperature and humidity measurement. Product information on demand (models HPP, ICH, ICH L, ICH C, IPPplus and ICOmed) B04714
External measuring instrument with additional measuring head for temperature and humidity measurement. Product information on demand B04714