Options / Accessories

for IVF module for ICOmed

Ideally suited for in vitro fertilisation. Controlled CO2 and O2 concentration as well as gentle cultivation in Petri dishes in separate slide-in units.

Option

Model sizes
Voltage 115 V, 50/60 Hz (no extra cost) X2 X2 X2 X2
Battery-buffered ControlCOCKPIT. Uninterrupted supply for the entire display unit (ControlCOCKPIT) and therefore complete documentation of all parameteres even when there is a power failure. The CO2 parameter is continously regulated. C2 C2 C2 C2
Two gas connections with quick release connectors, automatic switch-over of gas cylinders; incl. two CO2 connection sets: hose with coupling and clamp T1 T1 T1 T1
Electropolished interior T2 T2 T2 T2
Active microprocessor control for humidifying and dehumidifying (40 - 97 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times while avoiding condensate formation. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter (combination of options K7 and F7 for ICO50med only possible with accessory B44128) K7 K7 K7 K7
Control of oxygen concentration by N2 inlet; adjustment range 1 % up to 20 % O2; setting accuracy 0.1 % (requires option K7) T6 T6 T6 T6
Peltier cooling unit enables a working temperature of 37 °C even at higher ambient temperatures of up to 35 °C K5 K5 K5 K5
Capacitive humidity sensor for measuring and displaying the relative humidity K6 K6 K6 K6
Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by silicone stopper, at the back, centre right; not available for ICO50med with active humidity control (option K7, only in combination with accessory B44128) or humidity display (option K6) F7 F7 F7 F7
Heated inner door with partitioned glass doors (cannot be used in connection with option B8); size105/150/240 has 2/3/4 partitioned glass doors K4 K4 K4
4 – 20 mA current loop interface O2 controller, actual value (0 – 25 % O2 4 – 20 mA) V1 V1 V1 V1
4 – 20 mA current loop interface CO2 controller, actual value (0 – 25 % CO2 4 – 20 mA) V9 V9 V9 V9
4 – 20 mA current loop interface Humidity controller, actual value (0 – 100 % rh 4 – 20 mA) V7 V7 V7 V7
4 – 20 mA current loop interface Temperature controller, actual value (0 to +70 °C 4 – 20 mA) V3 V3 V3 V3
Works calibration certificate for 5 %, 7 % and 10 % CO2 (measured at +37 °C) special works calibration certificates upon request D00106 D00106 D00106 D00106
Works calibration certificate for one (freely selectable) temperature, humidity, CO2 and O2 value according to customer specification (requires T6) D00143 D00143 D00143 D00143
Works calibration certificate for one (freely selectable) temperature, humidity and CO2 value according to customer specification (requires K7) D00131 D00131 D00131 D00131
Start-up of ICOmed incubators and brief training (D, A, CH only), through Memmert service, not subject to discount K9 K9 K9 K9
Door hinged on the left B8 B8 B8 B8
Potential-free contact (24V/2A) with socket 3 pin,
according to NAMUR NE 28 for external monitoring
(setpoint of temperature and CO<sub>2</sub> is reached) Potential-free contact (24V/2A) with socket 3 pin, according to NAMUR NE 28 for external monitoring (setpoint of temperature and CO2 is reached) H5 H5 H5 H5
Potential-free contact for combination error message (e.g. supply failure, sensor fault, fuse) Potential-free contact for combination error message (e.g. supply failure, sensor fault, fuse) H6 H6 H6 H6
MobileALERT for up to 4 alarm notifications; standard: temperature and CO2 alarm, additionally humidity alarm (when equipped with option K7) and O2 alarm (when equipped with option T6) C4 C4 C4 C4
MobileALERT, notification by SMS in case of any error or alarm of the device. Requires option H6 “floating contact for alarm” C3 C3 C3 C3

Accessory

Model sizes
Additional perforated stainless steel shelf E35160 E37418 E37418 E35158
Additional water dish (not applicable with option K7) B38737 B38737 B38737 B38737
Central water supply with filter cartridges for connection to the domestic water supply, only in combination with option K7. Product information on demand ZWVR6 ZWVR6 ZWVR6 ZWVR6
Central water supply, without filter cartridges for connection to the domestic water supply (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7) only in combination with option K7. Product information on demand ZWVR7 ZWVR7 ZWVR7 ZWVR7
HEPA-filter for chamber according to EN 1822, packed in sterile condition, incl. fixing unit B38739 B38739 B38739 B38739
Guarantee extension by 1 year GA3Q5 GA3Q5 GA3Q5 GA3Q5
IVF module for ICO50med: patented, consisting of 6 slide-in units, a total of 12 special racks with indentations for 12 Petri dishes (60 mm diam.) resp. 24 Petri dishes (35 mm diam.), 2 racks with indentations for 3 special media tubes each; racks with indentations for 4-well dishes on demand; only for ICO50med with the options K7 and F7; works calibration certificate (measuring point chamber centre) at +37 °C, 5 %, 6 % and 7 % CO2 as well as 90 % rh; 5 % O2 for IVF unit equipped with option T6 B44128
IVF module for ICO105med: patented, consisting of 8 slide-in units, a total of 16 special racks with indentations for 16 Petri dishes (60 mm diam.) resp. 32 Petri dishes (35 mm diam.), 2 racks with indentations for 3 special media tubes each; racks with indentations for 4-well dishes on demand; only for ICO105med with the options K7 and F7; works calibration certificate (measuring point chamber centre) at +37 °C, 5 %, 6 % and 7 % CO2 as well as 90 % rh; 5 % O2 for IVF unit equipped with option T6 B42398
Holder for test tube (only in combination with IVF-module) E37069 E37069
Holder for Petri dishes square (only in combination with IVF-module) E37308 E37308
Holder for Petri dishes round (only in combination with IVF-module) E37026 E37026
Magnetic foil, can be labelled with a non-permanent board marker (only in combination with IVF-module) E07620 E36651
USB-Ethernet adapter E06192 E06192 E06192 E06192
Ethernet connection cable 5 m for computer interface E06189 E06189 E06189 E06189
USB User-ID stick (with User-ID licence): Oven-linked authorisation licence (User-ID-programme) on Memory-stick, prevents undesired manipulation by unauthorised third parties. When reordering please specify serial number (only for units with TwinDISPLAY) B33170 B33170 B33170 B33170
Set of height adjustable feet (4 pcs). Standard on models ICOmed B29768 B29768 B29768 B29768
Stacking set (4 pcs) for stacking of appliances of same size (not for models 160, 260, 450, 750, 1060, ICH110, ICH110L and ICH110C) Stacking set (4 pcs) for stacking of appliances of same size (not for models 160, 260, 450, 750, 1060, ICH110, ICH110L and ICH110C) B29744 B29744
Stacking set (consisting of stacking corners, one connecting plate for the rear, two wall brackets) for stacking two units B42114
Stacking set (consisting of stacking corners, one connecting plate for the rear, two wall brackets) for stacking of two units of same size (only in combination with subframe B33509 or B43598) B48129
Subframe (130 mm high); sizes 150/240: only in combination with the corresponding stacking sets for stacked appliances. B33507 B33508 B33508 B33509
Subframe (622 mm high) adjustable in height (sizes 150/240: should not be used for 2 stacked units) B33504 B33505 B33505 B33506
Subframe, on castors (height 120 mm; stainless steel, material 1.4301) B43598
CO2 pressure reducing valve DIN 8546, incl. Gas cylinder monitor E02087 E02087 E02087 E02087
N2 pressure reducing valve to DIN EN ISO 2503, incl. gas cylinder monitor (requires option T6) E06162 E06162 E06162 E06162
Celltron benchtop shaker (not subject to discount) – accessories upon request E06724 E06724 E06724
Integration of one additional unit (up to max. 31 units) into an already existent FDA-software licence FDAQ2 FDAQ2 FDAQ2 FDAQ2
IQ/OQ document with device-specific works test data for one freeselectable CO2 and temperature value, incl. temperature distribution survey at Memmert for 27 measuring points to DIN 12880:2007-05, PQ check list as support for validation by customer (models ICOmed). D38898 D38898 D38898 D38898
IQ/OQ document with device-specific works test data for one free-selectable CO2, humidity and temperature value, incl. temperature distribution survey at Memmert for 27 measuring points to DIN 12880:2007-05, PQ check list as support for validation by customer (models ICH C and ICOmed, on models ICOmed a free-selectable humidity value is only possible with option K7). D38897 D38897 D38897 D38897
Software conforming to FDA AtmoCONTROL. Meets the requirements for the use of electronically stored data sets and electronic signatures as laid down in Regulation 21 CFR Part 11 of the US Food and Drug Administration (FDA). Base licence for reading the data logging of one unit (only for units with TwinDISPLAY). Respective IQ/OQ documents available in German and English language (without surcharge) FDAQ1 FDAQ1 FDAQ1 FDAQ1
IQ document with device-specific works test data, OQ/PQ check list as support for validation by customer D00124 D00124 D00124 D00124
External measuring instrument with sensors with additional measuring head for temperature and humidity measurement. Product information on demand (models HPP, ICH, ICH L, ICH C, IPPplus and ICOmed) B04714 B04714 B04714 B04714
External measuring instrument with additional measuring head for temperature and humidity measurement. Product information on demand B04714 B04714 B04714 B04714