CO2 Incubator (med) INCO108med

Protection for in vitro growing cell cultures, tissue cultures and bacteria cultures in a controlled and physiologically ideal environment.

INCO108med
Medical Device CE 0197

CO2 Incubator (med) INCO108med

Standard delivery CO2 Incubator (med)

On this page, you can find all the essential technical data on the Memmert CO2 incubator INCOmed. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.

Properties

Temperature

Working temperature range min. 8°C above ambient up to +50°C
Temperature 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value

Humidity

Humidity integrated humidity limit control (88-97 % rh) with digital display of relative humidity - resolution of display 0.5 %, setting accuracy 1 %

CO2

CO2 CO2-supply via sterile filter
CO2 intrinsic adjustment of altitude above sea level
CO2 CO2-supply interruption upon door opening
CO2 digital electronic CO2-control with automatic zeroing (infrared measuring system) - adjustment range 0 - 10 % - resolution 0.1 %

Control of standard components

Controller digital display of setpoit and actual values (LED) for temperature (resolution 0.1°C) and CO2 concentration (0.1 %)
Timer digital 7-day programme timer with real time clock, precise minute setting

Control technology

Display digital display of all set parameters, such a temperature, weekdays, time, CO2, humidity and set-up values - language to be chosen via set-up
Controller Electronic microprocessor temperature controller with auto-diagnostic system
Calibration three freely selectable temperature values, 2-point-calibration for humidity: 20 and 90 %, 3-point calibration for CO2: 5 %, 7 %, 10 %

Communication

Documentation programme stored in case of power failure

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system integral fault diagnostics for temperature, CO2 and humidity limit control
Alarm Acoustic alarm: Over- and undertemperature, over-CO2 and empty gas cylinder, open door, underhumidity and empty water tank (with optional humidity module)

Heating concept

6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Ventilation uniform atmosphere and temperature distribution owing to enclosed non-turbulent ventilation system in working chamber

Standard equipment

Works calibration certificate incl. works calibration certificate for +37°C
Scope of delivery chip-card (STERICard) for sterilisation of working chamber with fixed values (4 hours/160°C), without removal of sensors and mountings
Door fully insulated stainless steel door with 2-point locking (compression door lock), lockable
Door inner glass door with opening (8 mm Ø) to take gas sample
Internals 2 perforated stainless steel shelves and 1 stainless steel water tray

Stainless steel interior

Interior easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing, material 1.4301 (ASTM 304), hermetically welded
Volume 108 l
Dimensions W x H x D in mm w(A) x h(B) x d(C): 560 x 480 x 400 mm
Max. number of grids/shelves 4
Max. loading per grid/shelf 0 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 710 x 778 x 550 mm
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1000 W

Packing/shipping data

the appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton B x H x T: 830 x 1060 x 800 mm
Net weight approx. 70 kg
Gross weight carton approx. 95 kg

More product series of Incubators

CO<sub>2</sub> Incubator

CO2 Incubator

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Cooled storage incubator

Cooled storage incubator

Sample breeding and storage in microbiology, food, cosmetics or pharmaceutics: The cooled storage incubator with Peltier elements lowers operating costs as well as CO2 emissions.

Compressor-cooled incubator

Compressor-cooled incubator

Do you require different ramps for incubating, breeding or storing? Do you need a larger chamber volume and at the same time quick temperature changes?

Peltier-cooled incubator

Peltier-cooled incubator

Microbiology, zoology, food, cosmetics or pharma industry: the energy-saving cooled incubator with Peltier elements heats up and cools down seamlessly in one system.

Incubator

Incubator

The incubator I is at home everywhere in the world of research, medicine, pharmaceutics and food analytics, as well as food chemistry.

IVF module for INCOmed

IVF module for INCOmed

Ideally suited for in vitro fertilisation. Controlled CO2 and O2 concentration as well as gentle cultivation in Petri dishes in separate slide-in units.

More product categories

Heating / drying ovens

Climate chambers

Water and oilbaths