Temperature

Setting temperature range +10 to +95 °C and boiling stage
Working temperature range in °C min. 5 above ambient up to +95 °C with additional boiling mode
resolution of display and setting accuracy 0,1°C up to +99,9°C / 1°C from 100°C

Control of standard components

Controller digital display (LED) of all set parameters, such as temperature and alarm values (0,1°C resolution) as well as time values
Timer digital timer from 1 min. up to 999 hours for: ON (continuous operation), DELAYED ON, HOLD or HOLD set-temperature dependent with guaranteed holding time
Calibration on controller

Safety

Temperature sensor 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Temperature control mechanical temperature limiter TB protection class 1 switching the heating off at approx. 30°C above max. temperature of the bath
Temperature control independently working, electronic overtemperature controller TWW protection class 3.1 or overtemperature limiter TWB protection class 2, adjustable in set-up menu by customer
Temperature control in case of overtemperature due to failure, the heating is switched off at approx. 10°C above the set temperature (fixed value)
Autodiagnostic system fuzzy-supported PID microprocessor controller with integrated autodiagnostic system with fault indication
Alarm visual and acoustic alarm at programme end and as input acknowledgement as well as in case of low liquid level, heating is switched off automatically

Heating concept

Heating Baths corrosion-proof large-area heating on three sides

Stainless steel interior

Dimensions W x H x D in mm w(A) x h(C) x d(B): 240 x 140 x 210 mm
Interior easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing, material no. 1.4301 (ASTM 304), laser-welded
Volume 7 l
Liquid level min. (H) 97 mm
Liquid level max. (I) 120 mm

Textured stainless steel casing

Dimensions w(D) x h(G) x d(E): 468 x 337 x 356 mm

Electrical data

Voltage 230 V, 50/60 Hz
Electrical load approx. 1200 W (during heating)

Ambient conditions

Installation The vent openings in the left and back side must remain unobstructed. Minimum wall spacing on all sides is 80 mm. The minimum spacing from the top of the bath to the next ceiling is 750 mm.
Ambient temperature +5 °C to +40 °C
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 570 x 450 x 420 mm
Net weight approx. 11 kg
Gross weight carton approx. 16 kg

Standard units are safety-approved and bear the test marks

CE
EAC Eurasian Conformity
VDE-EMC electromagnetic compatibility