On this page, you can find all the essential technical data on the Memmert water bath. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at sales@memmert.com.
Setting temperature range | +10 to +95 °C and boiling stage |
---|---|
Working temperature range in °C | min. 5 above ambient up to +95 °C with additional boiling mode |
Temperature sensor | 1 Pt100 sensor class A in 4-wire-circuit |
---|---|
Timer | integrated digital timer from 1 min. to 99,59 hours for: ON continuous operation WAIT (delayed on for continuous and limited timed operation) HOLD |
Controller | digital display of set and actual temperature and of (remaining) programme time |
---|
Temperature control | mechanical temperature limiter TB protection class 1 switching the heating off at approx. 30°C above max. temperature of the bath |
---|---|
Temperature control | in case of overtemperature due to failure, the heating is switched off at approx. 10°C above the set temperature (fixed value) |
Autodiagnostic system | microprocessor PID-temperature controller with integrated autodiagnostic system with fault indicator |
Heating Baths | corrosion-proof large-area heating on three sides |
---|
Dimensions W x H x D in mm | w(A) x h(C) x d(B): 590 x 140 x 350 mm |
---|---|
Interior | easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing, material no. 1.4301 (ASTM 304), laser-welded |
Volume | 29 l |
Liquid level min. | 105 mm |
Liquid level max. | 120 mm |
Dimensions | w(D) x h(G) x d(E): 818 x 343 x 516 mm |
---|
Voltage | 230 V, 50/60 Hz |
---|---|
Electrical load | approx. 2400 W (during heating) |
Installation | The vent openings in the left and back side must remain unobstructed. Minimum wall spacing on all sides is 80 mm. The minimum spacing from the top of the bath to the next ceiling is 750 mm. |
---|---|
Ambient temperature | +5 °C to +40 °C |
Humidity rh | max. 80 %, non-condensing |
Overvoltage category | II |
Pollution degree | 2 |
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | w x h x d: 910 x 610 x 400 mm |
Net weight | approx. 24 kg |
Gross weight carton | approx. 31 kg |