On this page, you can find all the essential technical data on the Memmert water bath. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at sales@memmert.com.
Temperature sensor | digital temperature sensor, mounted on the outside of the bath |
---|---|
Temperature | Setting Celsius / Fahrenheit |
Working temperature range | at least 5 above ambient to +100 (depending on ambient pressure) |
Setting temperature range | +10 to +100 °C |
Setting accuracy temperature | 0.1 °C |
Calibration | 2-point calibration |
---|---|
Controller | intuitive 3.5" colour graphic display with touch screen |
Controller | digital display of set and actual temperature and of (remaining) programme time |
Temperature control | two-stage safety system to prevent overtemperature |
---|---|
Temperature control | individually adjustable overtemperature alarm, visual and acoustic |
Temperature control | Independent temperature limiter to switch off the heating in case of failure |
Autodiagnostic system | microprocessor PID-temperature controller with integrated autodiagnostic system with fault indicator |
Heating | corrosion-proof large-area heating on the outside of the interior |
---|
Drain system | innovative drain system - easy to empty |
---|---|
Timer | digital timer from 1 min. up to 99 hours 59 min., visual and acoustic |
Interior | Deep-drawn stainless steel tray (stainless steel W.St. 1-4301), without corners and edges for easy cleaning incl. central drain (no disturbing installations) |
---|---|
Dimensions W x H x D in mm | w(A) x h(C) x d(B): 482 x 150 x 277 mm |
Volume | 17 l |
Liquid level min. | 20 mm |
Liquid level max. | 114 mm |
Dimensions | w(D) x h(E2) x d(F): 554 x 348 x 349 mm |
---|
Voltage Electrical load | 230 V, 50/60 Hz approx. 1420 W |
---|---|
Voltage Electrical load | 115 V, 50/60 Hz approx. 1420 W |
Installation | The vent openings in the left and back side must remain unobstructed. Minimum wall spacing on all sides is 80 mm. The minimum spacing from the top of the bath to the next ceiling is 750 mm. |
---|---|
Ambient temperature | +5 °C to +40 °C |
Humidity rh | max. 80 %, non-condensing |
Overvoltage category | II |
Pollution degree | 2 |
Transport information | The appliances must be transported upright |
---|---|
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton | w x h x d: 670 x 530 x 400 mm |
Net weight | approx. 9 kg |
Gross weight carton | approx. 14 kg |