Download Product Sheets (ZIP)
On this page, you can find all the essential technical data on the Environmental test chambers TESTA CT / TT. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Range of functions
| Adjustable parameters | Temperature |
|---|---|
| Adjustable device functions | Climate control regulation; Real-time display of process variables; Program execution; Alarm definition and much more |
| Control unit | Touch screen controller |
| Interfaces | Ethernet LAN |
Temperature
| Cooling system | Air cooled hermetic compressor group (low noise and high efficiency) with enforced ventilation and without CFC’s. |
|---|---|
| Temperature control system | By stainless steel electric heaters located in the air treatment tunnel |
| Temperature fluctuations with time | ± 0,1 to ± 0,3 K |
| Setting temperature range | -52 to +180 °C |
| Resolución del indicador valores nominales de la temperatura y valores reales | 0,1 °C |
Interior de acero inoxidable
| Medidas | An(A) x Al(B) x F(C): 984 x 910 x 1052 mm (F menos los 35 mm del ventilad) |
|---|---|
| Volumen | 967 l |
| N° máx. de inserciones | 10 |
| Máx. carga de la cámara del equipo: | 200 kg |
| Carga máx. por inserción | 50 kg |
Carcasa de acero estructural
| Medidas | An(D) x Al(E) x F(F): 1359 x 2010 x 1845 mm (F +56mm manilla de la puerta) |
|---|---|
| Instalación | Ruedas para desplazamientos con posibilidad de bloqueo, nivelable |
Datos eléctricos
| Potencia | 400 V, 3ph. w/o N 16A, 50 Hz / aprox. 22 W |
|---|
Condiciones ambientales
| Ambient conditions | Distance to neighbouring appliances / wall clearances: 1000 mm back; 500 mm lateral |
|---|
Datos de embalaje/envío
| Country of origin | Portugal |
|---|---|
| Información de transporte | ¡Los equipos deben transportarse en posición vertical! |
| Número estadístico de mercancía | 8419 8998 |
| Medidas aprox. incl. cartón | An x Al x F 1590 x 2190 x 2090 mm |
| Peso neto | aprox. 874 kg |
| Peso bruto cartón | aprox. 950 kg |
Safety
| Temperature monitoring | Safety thermostat with High / Low temperature configuration, with automatic stop of all thermic systems |
|---|---|
| Alarm | High / Low Temperature alarms |
| Temperature monitoring | 1 Pt100 sensor DIN class A; - 1 Pt100 sensor DIN class A, movable sensors for flexible placing inside chamber |
Standard equipment
| Standard option | Entryport Ø 80 mm, left; Air-cooled condenser (standard on -50 models); Door locking, mechanical |
|---|---|
| Door | Lockable door, double silicone door gaskets and anti-condensation heating frames |
| Additional accessories | Water conductivity measuring device; Documentation package |
| Inserciones | 2 bandeja(s) perforada(s) de acero inoxidable |
Las unidades estándar están aprobadas en materia de seguridad y llevan las marcas certificadoras