Download Product Sheets (ZIP)
On this page, you can find all the essential technical data on the ULT Freezer. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Range of functions
| Adjustable parameters | Temperature |
|---|---|
| Control unit | 7” touchscreen integrated in the outer door in ergonomic position |
Temperature
| Setting temperature range | -50 to -86 °C |
|---|---|
| Cooling system | Cooling system with two (2) cascade-connected compressors for reliable low-temperature performance. Refrigerants: First stage R455A / Second stage R170. Air-cooled condenser. |
| Résolution d'affichage des valeurs de la température de consigne et de la température réelle | 0,1 °C |
Safety
| Power failure | Backup battery for alarming and monitoring during power failure |
|---|---|
| Temperature monitoring | Temperature sensor Pt1000 |
| Alarme | Visuelle et sonore |
Communication
| Interface | Ethernet LAN, USB |
|---|
Standard equipment
| Standard option | Door-open detection; Entry port / pressure relief opening Ø 26 mm, lateral, Battery-backed alarm system |
|---|---|
| Door | Insulated stainless steel outer doors with mechanical door lock |
| Additional accessories | Standard interior configuration with 4 compartments. Consisting of 2 stainless steel inner doors and 3 height-adjustable shelves for flexible adaptation to different load sizes. |
Caisson intérieur en acier inoxydable
| Medidas | w(A) x h(B) x d(C): 600 x 1345 x 627 mm |
|---|---|
| Max. number of 2" / 50mm boxes | 384 |
| Volume | 506 l |
| Charge maximale de l'appareil: | 360 kg |
| Charge max. par clayette | 180 kg |
Caisson extérieur en acier inox structuré
| Mésures | w(D) x h(E) x d(F): 850 x 2000 x 888 mm |
|---|---|
| Installation | Sur roulettes à frein intégré |
Données sur l'emballage/l'expédition
| Customs tariff number | 8418 40 80 |
|---|---|
| Pays d'origine | Netherlands |
| Information du transport | Les appareils doivent être transportés en position verticale! |
| Dimensions env., carton inclus | Dimensions avec l'emballage (L x H x P): 1080 x 2220 x 1080 mm |
| Poids net | approximatif: 343 kg |
| Poids brut sous carton | approximatif: 430 kg |
Les unités standards ont reçu l’approbation de sécurité et contiennent les données de test