On this page, you can find all the essential technical data on the Environmental test chambers TESTA CT / TT. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Range of functions
| Adjustable parameters | Temperature; Relative humidity |
|---|---|
| Adjustable device functions | Climate control regulation; Real-time display of process variables; Program execution; Alarm definition and much more |
| Control unit | Touch screen controller |
| Interfaces | Ethernet LAN |
Temperature
| Setting temperature range | -75 to +180 °C |
|---|---|
| Cooling system | Water-cooled condensers are also available as standard in -75ºC models or an option for models with temperature cooling rate upgrades. |
| Temperature control system | By stainless steel electric heaters located in the air treatment tunnel |
| Temperature fluctuations with time | ± 0,1 to ± 0,3 K |
| Résolution d'affichage des valeurs de la température de consigne et de la température réelle | 0,1 °C |
Humidity
| Setting range humidity | 10 to 98 %rh |
|---|---|
| Humidification system | Humidity: Through thermostatic bath with dew point control. Drying: Through thermostatic bath with dew point control and additional dry coil |
| Accuracy | 0,1 % rh |
Safety
| Temperature monitoring | Safety thermostat with High / Low temperature configuration, with automatic stop of all thermic systems |
|---|---|
| Temperature monitoring | 1 Pt100 sensor DIN class A; 1 Pt100 sensor DIN class A, movable sensors for flexible placing inside chamber; Capacitive and psychrometric humidity sensor | ECP |
| Alarm | Automatic stop function in case of water failure, with indication on the controller: High / Low humidity alarms; High / Low Temperature alarms |
Standard equipment
| Standard option | Entryport Ø 80 mm, left; Water-cooled condenser (standard on models -75 and -40); Door locking, mechanical |
|---|---|
| Door | Lockable door, double silicone door gaskets and anti-condensation heating frames |
| Additional accessories | Water conductivity measuring device; Documentation package |
| Clayettes | 2 plateau(x) perforé(s) inox |
Caisson intérieur en acier inoxydable
| Mésures | L(A) x H(B) x P(C): 634 x 715 x 600 mm (P moins 35mm pour ventilateur) |
|---|---|
| Volume | 272 l |
| Nombre max. clayettes | 10 |
| Charge maximale de l'appareil: | 120 kg |
| Charge max. par clayette | 25 kg |
Caisson extérieur en acier inox structuré
| Mésures | L(D) x H(E) x P(F): 1009 x 1840 x 1340 mm (P +56mm hors poignée) |
|---|---|
| Installation | Sur roulettes à frein intégré, support de mise à niveau |
Données électriques
| Puissance | 400 V, 3ph. w/o N 16A, 50 Hz / environ 11 W |
|---|
Conditions d'environnement
| Ambient conditions | Distance to neighbouring appliances / wall clearances: 1000 mm back; 500 mm lateral |
|---|
Données sur l'emballage/l'expédition
| Country of origin | Portugal |
|---|---|
| Information du transport | Les appareils doivent être transportés en position verticale! |
| Tarif douanier commun | 8419 8998 |
| Dimensions env., carton inclus | Dimensions avec l'emballage (L x H x P): 1110 x 2090 x 1520 mm |
| Poids net | approximatif: 540 kg |
| Poids brut sous carton | approximatif: 665 kg |
Les unités standards ont reçu l’approbation de sécurité et contiennent les données de test