Technical Specification

Humidity chamber HCP50

Active humidity control creates a controlled environment for accelerated life tests and 85/85 tests.

On this page, you can find all the essential technical data on the Memmert humidity chamber HCP. Our customer relations team will be pleased to help if you want further information.

Temperature

Working-temperature range +7 above ambient temperature up to +90 °C
Setting temperature range +18 to +90 °C
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity

Humidity Control Active microprocessor control for humidifying and dehumidifying (20 – 95 % rh), incl. digital indication and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times. Humidity supply with water (only for demineralised water with a conductivity of 5 to 10 μS/cm and a pH value between 5 and 7; from an external tank) by a self-priming pump; integral bacteria block by generating hot steam, dehumidifying via sterile filter
Setting range active humidity control 20 to 95 % rh and rh-Off
Setting accuracy humidity 0.5 % rh

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable values each, temperature and humidity

Communication

Interface Ethernet LAN,
USB
Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic

Heating concept

6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid condensation

Standard equipment

Works calibration certificate Works calibration certificate (measuring point chamber centre) at 60 °C with 75 % rh
Door Additional heated inner glass door
Door fully insulated stainless steel door with 2-point locking (compression door lock)
Entry port Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone stopper, at the back, centre left
Internals 1 perforated stainless steel shelf/shelves
Standard accessories Water tank including connection hose

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 400 x 425 x 330 mm (d less 35 mm for fan)
Interior W.-St. 1.4301 (ASTM 304)
Volume 56 l
Max. number of internals 5
Max. loading of chamber 75 kg
Max. loading per internal 15 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 559 x 795 x 521 mm (d +56mm door handle)

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1520 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1520 W

Ambient conditions

Ambient temperature 10 °C to 30 °C
Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 730 x 950 x 640 mm
Net weight approx. 55 kg
Gross weight carton approx. 74 kg

Standard units are safety-approved and bear the test marks

CE
VDE
cMET

Let us help you find your Memmert

Send inquiry

Always ahead for you.

2026 – A year full of innovations.