Technical Specification

Peltier-cooled incubator IPP410eco

Sets eco standards for cultivation below room temperature: unmatched energy efficiency, best values for heat-up, cool-down and recovery times.

With the help of our model selection, dimensional model sketches and extensive technical data for download, you will find your perfect Peltier-cooled incubator. Flexibility and technical features of our appliances meet all possible needs. Put us to the test!

Temperature

Setting temperature range 0 to +70 °C
Working temperature range with light: +10 to +40 °C
Working temperature range without light: from 0 (at least 20 below ambient temperature) to +70°C
Setting accuracy temperature 0.1 °C
Temperature sensor 1 Pt100 sensor DIN class A in 4-wire-circuit

Control technology

Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
ControlCOCKPIT SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime

Ventilation

Convection forced ventilation by Peltier fan

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).

Safety

Temperature control adjustable electronic overtemperature monitor and mechanical temperature limiter
Autodiagnostic system for fault analysis

Heating concept

Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)

Standard equipment

Works calibration certificate for +10 °C, +25 °C and +40 °C
Door fully insulated stainless steel door with 2-point locking (compression door lock)
Door inner glass door
Internals 2 stainless steel grid(s), electropolished

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 640 x 1200 x 500 mm (d less 32 mm for fan - Peltier)
Volume 384 l
Max. number of internals 14
Max. loading of chamber 200 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 824 x 1720 x 655 mm (d +56mm door handle & +111mm Peltier element)
Installation on lockable castors
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 600 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 600 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 930 x 1930 x 930 mm
Net weight approx. 139 kg
Gross weight carton approx. 190 kg

Standard units are safety-approved and bear the test marks

CE

Let us help you find your Memmert

Send inquiry

Always ahead for you.

2026 – A year full of innovations.