Download Product Sheets (ZIP)
On this page, you can find all the essential technical data on the ULT Freezer. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Range of functions
| Adjustable parameters | Temperature |
|---|---|
| Control unit | 7” touchscreen integrated in the outer door in ergonomic position |
Comunicación
| Interface | Ethernet LAN, USB |
|---|
Interior de acero inoxidable
| Dimensions | w(A) x h(B) x d(C): 450 x 1111 x 489 mm |
|---|---|
| Max. number of 2" / 50mm boxes | 162 |
| Volumen | 244 l |
| Máx. carga de la cámara del equipo: | 280 kg |
| Carga máx. por inserción | 140 kg |
Carcasa de acero estructural
| Medidas | w(D) x h(E) x d(F): 700 x 1832 x 738 mm |
|---|---|
| Instalación | Ruedas para desplazamientos con posibilidad de bloqueo |
Datos de embalaje/envío
| Customs tariff number | 8418 40 20 |
|---|---|
| País natal | Netherlands |
| Información de transporte | ¡Los equipos deben transportarse en posición vertical! |
| Medidas aprox. incl. cartón | An x Al x F 900 x 2050 x 900 mm |
| Peso neto | aprox. 245 kg |
| Peso bruto cartón | aprox. 310 kg |
Temperature
| Setting temperature range | -50 to -86 °C |
|---|---|
| Cooling system | Cooling system with two (2) cascade-connected inverter compressors for precise temperature control and high energy efficiency. Refrigerants: First stage R290 / Second stage R170. Air-cooled condenser. |
| Resolución del indicador valores nominales de la temperatura y valores reales | 0,1 °C |
Safety
| Power failure | Backup battery for alarming and monitoring during power failure |
|---|---|
| Temperature monitoring | Temperature sensor Pt1000 |
| Alarma | óptica y acústica |
Standard equipment
| Standard option | Door-open detection; Entry port / pressure relief opening Ø 26 mm, lateral, Battery-backed alarm system |
|---|---|
| Door | Insulated stainless steel outer doors with mechanical door lock |
| Additional accessories | Standard interior configuration with 4 compartments. Consisting of 2 stainless steel inner doors and 3 height-adjustable shelves for flexible adaptation to different load sizes. |
Las unidades estándar están aprobadas en materia de seguridad y llevan las marcas certificadoras