Technical Specification

HPP1060

New

Temperature

Setting temperature range without light, with humidity: +5 to +70 °C
Setting temperature range without light, without humidity: 0 to +70 °C
Working temperature range without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C
Working temperature range without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C

Humidity

Setting range humidity 10 - 90 % rh

Control technology

Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %

Standard equipment

Works calibration certificate for +25 °C/40 % rh and +40 °C/75 % rh
Door fully insulated stainless steel doors with 2-point locking (compression door lock)

Textured stainless steel casing

Installation on lockable castors

Electrical data

Voltage Electrical load 115 V, 50/60 Hz
approx. 1600 W

Packing/shipping data

Dimensions approx incl. carton w x h x d: 1370 x 1970 x 1300 mm
Gross weight carton approx. 424 kg

Standard units are safety-approved and bear the test marks

CE

Let us help you find your Memmert