Constant Climate Chamber m360 HPP1400Ceco

Sustainable climate simulation with maximum efficiency

HPP1400Ceco

Constant Climate Chamber m360 HPP1400Ceco

Standard delivery Constant Climate Chamber m360

On this page, you can find all the essential technical data on the m360 cConstant climate chamber HPPCeco. Our customer relations team will be pleased to help if you want further information. Please contact our technical specialists at m360@memmert.com.

Temperature

Setting temperature range without light, with humidity: +15 to +60 °C
Working temperature range without light, with humidity: +15 (at least 10 below ambient temperature) to +60 °C
Working temperature range without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity

Setting range humidity 10 - 80 % rh
Humidity humidity supply with distilled water from external tank by self-priming pump
Humidification humidification by hot steam generator
Setting accuracy humidity 0.5 % rh

CO2

Temporal fluctuation ± 0,4 % CO2
Setting accuracy 0,1 % CO2
Setting range 0 to 20 % CO2

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable values each, temperature/humidity/CO2

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic

Heating concept

Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)

Standard equipment

Standard works calibration certificate +25 °C / 60 % rh, +40 °C / 75 % rh, 5 % CO2 at +40 °C / 75 %rh
Door Door-open-recognition
Door Stainless steel doors with glass sectors, fully heated inner glass panes integrated in the full-sight glass door with 2-point locking (compression door lock)
Standard equipment CO2 connection set: hose with coupling and clamp
Internals 4 stainless steel grid(s), electropolished
Standard accessories Water tank including connection hose

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 1250 x 1450 x 750 mm (d less 32 mm for fan - Peltier)
Volume 1360 l
Max. number of internals 28
Max. loading of chamber 250 kg
Max. loading per internal 30 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 1435 x 1913 x 905 mm (d +56mm door handle & +111mm Peltier element)
Installation on lockable castors, adjustable in height

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 1400 W
Voltage Electrical load 115 V, 50/60 Hz
approx. 1400 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. wooden box w x h x d: 1560 x 2200 x 1190 mm
Dimensions approx incl. carton w x h x d: 1560 x 2200 x 1190 mm
Net weight approx. 347 kg
Gross weight wooden box approx. 525 kg

Standard units are safety-approved and bear the test marks