The indispensable safety feature for this hot air steriliser: Setpoint Wait. This means that the programme for sterilisation only starts when the set temperature has been reached. This feature can also be used with freely positionable Pt100 temperature sensors. Here the sterilisation time only begins when the set temperature has been reached at all measurement points, and reliable sterilisation is guaranteed at all times.

On this page, you can find all the essential technical data on the Memmert hot air steriliser. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].

Temperature

Setting temperature range +20 to +250 °C
Setting accuracy temperature up to 99.9 °C: 0.1 / from 100 °C: 0.5
Working temperature range at least 10 above ambient temperature to +250°C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime

Ventilation

Fan forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
Vent vent connection with restrictor flap

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Temperature control overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system for fault analysis
Alarm visual and acoustic

Standard equipment

Works calibration certificate Calibration at +160°C
Door fully insulated stainless steel doors with 2-point locking (compression door lock), lockable
Internals 2 stainless steel grid(s), electropolished

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 1040 x 720 x 600 mm (d less 39 mm for fan)
Interior easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume 449 l
Max. number of internals 8
Max. loading of chamber 300 kg
Max. loading per internal 30 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 1224 x 1247 x 784 mm (d +56mm door handle)
Installation on lockable castors
Housing rear zinc-plated steel

Electrical data

Voltage Electrical load 400 V and 3x 230 V w/o neutral, 50/60 Hz approx. 5800 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installation max. 2,000 m above sea level
Ambient temperature +5 °C to +40 °C
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 1330 x 1440 x 1050 mm
Net weight approx. 161 kg
Gross weight carton approx. 227 kg

Standard units are safety-approved and bear the test marks

CE LGA
VDE
UL

Let us help you find your Memmert

Send inquiry

Intended use as medical device

Memmert is bringing medical devices of class I according to EU Directive (EU) 2017/745 into circulation.

Type Intended use as medical device Classification according to (EU) 2017/745
UNmplus, UFmplus The appliance may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm. I
UNm, UFm, INm, IFm The appliance may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm. I
INmplus, IFmplus Implus may be used for temperature control of rinsing and infusion solutions and contrast agents. I
IFbw The appliance is used to heat non-sterile cloths and covers. Any other use is improper
and may cause damage or danger.
I

Memmert is bringing medical devices of class IIa and IIb into circulation according to MDD 93/42/EEC until 31.12.2028 according to the transitional provisions (EU) 2023/607 by the (EU) 2017/745 article 120 (2).

Type Intended use as medical device Classification according to 93/42/EWG
SNplus, SFplus The product is intended for the sterilisation of medical devices with dry heat at atmospheric pressure. IIb
SN, SF The product is intended for the sterilisation of medical devices with dry heat at atmospheric pressure. IIb
ICOmed The CO2 incubator ICOmed is used to generate and maintain constant ambient conditions for the in-vitro fertilisation (IVF) application field, especially for the incubation of oocytes, spermatozoa and zygotes in special culture dishes for IVF application as well as for gene expression and the biosynthesis of RNA and proteins. IIa

Download as pdf

Always ahead for you.

2026 – A year full of innovations.