Download Product Sheets (ZIP)
On this page, you can find all the essential technical data on the Memmert ICO CO2 Incubator. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Range of functions
| Calibrate | Three freely selectable values: temperature, humidity, and CO2 |
|---|---|
| Interfaces | USB-C; Ethernet LAN |
| Timer | Timer - Process time starts immediately when the programme starts; Timer SetpointWAIT - Process time only starts after all setpoint values have been reached; Timer on Event - Process time starts at a defined point in time |
| Language settings | English (default), German, French, Spanish, Portuguese, Chinese, Arabic, Hindi, Polish, Czech, Hungarian, Russian, Italian, Serbian |
| Appliance functions | Date and time (time zones, summer/winter time); Display brightness; DistanceView (1–4 freely selectable parameters, activation time adjustable); Network settings (IP, subnet, DHCP); Language and volume settings (buzzer/key tone); Power failure behaviour; Timer functions |
| Adjustable parameters | Temperature; CO2; Relative humidity (requires option K7); O2 (requires option T6) |
| Control unit | 6.86" IPS touch display (480 × 1280 pixels) with multi-touch function for graphical display of actual/setpoint values and parameter curves. |
O2 (only with T6 option)
| Control system | Anaerobic O2 control for targeted oxygen reduction via nitrogen inlet. Zirconium oxide sensor with integrated monitoring and error display. Filtered N₂ supply for safe process conditions. |
|---|---|
| Adjustment precision | 0,1 % |
| O2 - Setting range | 1 to 20 % |
Temperature
| Temperature control system | Large-area multi-zone multifunctional all-round heating with additional door, seal and rear wall heating |
|---|---|
| Setting temperature range | +20 to +50 °C |
| Gamme des températures utiles | 5 °C au-dessus de la température ambiante à +50 °C Programme de stérilisation incorporé: 60 minutes à 180 °C (les sondes peuvent rester en place) |
| Résolution d'affichage des valeurs de la température de consigne et de la température réelle | 0,1 °C |
| Ecart d ́homogéniété | à +37 °C +/- 0.3 K |
| Ecart de justesse de température | à 37 °C +/- 0.1 K |
Humidity
| Humidification system | Passive humidification by means of integrated water tank with condensate return via magnetically guided sealing lip. Humidity limitation by Peltier element. Active closed-loop humidity control (option K7) with capacitive sensor for precise measurement and complete documentation. Humidification via self-priming micro-dosing pump with hot steam generation, dehumidification via powerful diaphragm pump. |
|---|---|
| Accuracy | 0,1% rh |
| Setting range | 40 to 98 and rh-Off |
CO2
| Control system | Precise CO2 control with dual-beam NDIR technology, integrated monitoring and pressure-compensated measurement. Filtered CO2 supply for high process reliability. |
|---|---|
| Temporal fluctuation | ± 0,2 % |
| Setting range | 0 to 20 % |
Sécurité
| Power failure | Manual mode: Power failure less than 24 hours: Active process is terminated, parameters remain stored. Power failure longer than 24 hours: Operation continues with previously set parameters. Program mode (power failure >1 hour): Failure less than 1 hour: Programme continues with remaining runtime. Failure longer than 1 hour: Programme is stopped; query for continuation required |
|---|---|
| Temperature monitoring | Dual Pt1000 sensor, DIN Class A, for mutual monitoring and automatic takeover in the event of a fault |
| Alarme | Visuelle et sonore |
Standard equipment
| Standard option | Door-open detection | V5 |
|---|---|
| Factory calibration certificate III (Option T6) | +37 °C, 5 %CO2, 90 %RH, 10 %O2 |
| Factory calibration certificate II (Option K7) | +37 °C, 5 %CO2, 90 %RH |
| Factory calibration certificate I (Standard) | +37 °C, 5 %CO2 |
| Additional accessories | CO2 connection set, hose with connection coupling and hose clamp; Water dish (Basic equipment); Water tank including connection hose (requires Option K7); Preset sterilisation programme: 60 minutes at +180 °C (without removing the sensors) |
| Porte | Porte en acier inoxydable entièrement isolée avec verrouillage 2 points (fermeture de la porte par compression) |
| Porte | Porte intérieure en verre avec ouverture (Ø 8 mm) pour prélèvement des échantillons gazeux |
| Clayettes | 1 plateau(x) perforé(s) inox |
Textured stainless steel casing
| Mésures | w(D) x h(E) x d(F): 548 x 628 x 519 mm |
|---|---|
| Installation | 4 adjustable appliance feet |
Caisson intérieur en acier inoxydable
| Medidas | w(A) x h(B) x d(C): 400 x 425 x 330 mm |
|---|---|
| Volume | 56 l |
| Nombre max. clayettes | 4 |
| Charge maximale de l'appareil: | 50 kg |
| Charge max. par clayette | 15 kg |
Données électriques
| Tension/Charge électrique | 230 V, 50/60 Hz environ 1200 W |
|---|---|
| Tension/Charge électrique | 115 V, 50/60 Hz environ 1200 W |
Conditions d'environnement
| Installation | Il convient de prévoir un écart de 15 cm minimum entre le mur et le panneau arrière de l'appareil. L'écart ne doit pas être inférieur à 20 cm avec le plafond et à 5 cm sur les côtés par rapport au mur ou à un autre appareil |
|---|---|
| Température ambiante | 10 °C à 35 °C |
| Altitude d'installation | max. 2000 m au-dessus du niveau moyen de la mer |
| Classe de surtension | II |
| Niveau de pollution | 2 |
Données sur l'emballage/l'expédition
| Information du transport | Les appareils doivent être transportés en position verticale! |
|---|---|
| Tarif douanier commun | 8419 8998 |
| Pays d'origine | Allemagne |
| Numéro d'enregistrement DEEE | DE 66812464 |
| Dimensions env., carton inclus | Dimensions avec l'emballage (L x H x P): 655 x 893 x 755 mm |
| Poids net | approximatif: 53 kg |
| Poids brut sous carton | approximatif: 66 kg |
Les unités standards ont reçu l’approbation de sécurité et contiennent les données de test