Download Product Sheets (ZIP)
On this page, you can find all the essential technical data on the Memmert ICO CO2 Incubator. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at [email protected].
Range of functions
| Calibrate | Three freely selectable values: temperature, humidity, and CO2 |
|---|---|
| Interfaces | USB-C; Ethernet LAN |
| Timer | Timer - Process time starts immediately when the programme starts; Timer SetpointWAIT - Process time only starts after all setpoint values have been reached; Timer on Event - Process time starts at a defined point in time |
| Language settings | English (default), German, French, Spanish, Portuguese, Chinese, Arabic, Hindi, Polish, Czech, Hungarian, Russian, Italian, Serbian |
| Appliance functions | Date and time (time zones, summer/winter time); Display brightness; DistanceView (1–4 freely selectable parameters, activation time adjustable); Network settings (IP, subnet, DHCP); Language and volume settings (buzzer/key tone); Power failure behaviour; Timer functions |
| Adjustable parameters | Temperature; CO2; Relative humidity (requires option K7); O2 (requires option T6) |
| Control unit | 6.86" IPS touch display (480 × 1280 pixels) with multi-touch function for graphical display of actual/setpoint values and parameter curves. |
O2 (only with T6 option)
| Control system | Anaerobic O2 control for targeted oxygen reduction via nitrogen inlet. Zirconium oxide sensor with integrated monitoring and error display. Filtered N₂ supply for safe process conditions. |
|---|---|
| Adjustment precision | 0,1 % |
| O2 - Setting range | 1 to 20 % |
Temperature
| Temperature control system | Large-area multi-zone multifunctional all-round heating with additional door, seal and rear wall heating |
|---|---|
| Setting temperature range | +20 to +50 °C |
| Working-temperature range | 5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors) |
| Setting accuracy temperature | 0.1 °C |
| Temperature variation in chamber | at + 37 °C +/- 0.3 K |
| Temperature fluctuations with time | at 37 °C +/- 0.1 K |
Humidity
| Humidification system | Passive humidification by means of integrated water tank with condensate return via magnetically guided sealing lip. Humidity limitation by Peltier element. Active closed-loop humidity control (option K7) with capacitive sensor for precise measurement and complete documentation. Humidification via self-priming micro-dosing pump with hot steam generation, dehumidification via powerful diaphragm pump. |
|---|---|
| Accuracy | 0,1% rh |
| Setting range | 40 to 98 and rh-Off |
CO2
| Control system | Precise CO2 control with dual-beam NDIR technology, integrated monitoring and pressure-compensated measurement. Filtered CO2 supply for high process reliability. |
|---|---|
| Temporal fluctuation | ± 0,2 % |
| Setting range | 0 to 20 % |
Safety
| Power failure | Manual mode: Power failure less than 24 hours: Active process is terminated, parameters remain stored. Power failure longer than 24 hours: Operation continues with previously set parameters. Program mode (power failure >1 hour): Failure less than 1 hour: Programme continues with remaining runtime. Failure longer than 1 hour: Programme is stopped; query for continuation required |
|---|---|
| Temperature monitoring | Dual Pt1000 sensor, DIN Class A, for mutual monitoring and automatic takeover in the event of a fault |
| Alarm | visual and acoustic |
Standard equipment
| Standard option | Door-open detection | V5 |
|---|---|
| Factory calibration certificate III (Option T6) | +37 °C, 5 %CO2, 90 %RH, 10 %O2 |
| Factory calibration certificate II (Option K7) | +37 °C, 5 %CO2, 90 %RH |
| Factory calibration certificate I (Standard) | +37 °C, 5 %CO2 |
| Additional accessories | CO2 connection set, hose with connection coupling and hose clamp; Water dish (Basic equipment); Water tank including connection hose (requires Option K7); Preset sterilisation programme: 60 minutes at +180 °C (without removing the sensors) |
| Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
| Door | inner glass door with opening (8 mm Ø) to take gas sample |
| Internals | 1 perforated stainless steel shelf/shelves |
Stainless steel interior
| Dimensions | w(A) x h(B) x d(C): 400 x 425 x 330 mm |
|---|---|
| Volume | 56 l |
| Max. number of internals | 4 |
| Max. loading of chamber | 50 kg |
| Max. loading per internal | 15 kg |
Textured stainless steel casing
| Dimensions | w(D) x h(E) x d(F): 548 x 628 x 519 mm |
|---|---|
| Installation | 4 adjustable appliance feet |
Electrical data
| Voltage Electrical load | 230 V, 50/60 Hz approx. 1200 W |
|---|---|
| Voltage Electrical load | 115 V, 50/60 Hz approx. 1200 W |
Ambient conditions
| Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
|---|---|
| Ambient temperature | 10 °C to 35 °C |
| Altitude of installation | max. 2,000 m above sea level |
| Overvoltage category | II |
| Pollution degree | 2 |
Packing/shipping data
| Transport information | The appliances must be transported upright |
|---|---|
| Customs tariff number | 8419 8998 |
| Country of origin | Federal Republic of Germany |
| WEEE-Reg.-No. | DE 66812464 |
| Dimensions approx incl. carton | w x h x d: 655 x 893 x 755 mm |
| Net weight | approx. 53 kg |
| Gross weight carton | approx. 66 kg |
Standard units are safety-approved and bear the test marks