Technical Specification

Cooled vacuum oven m360 Cooled vacuum oven m360 VO49cool

Innovative Vacuum Drying for Temperature-Sensitive Materials

On this page, you can find all the essential technical data on the m360 cooled vacuum oven VOcool. Our customer relations team will be pleased to help if you want further information. Please contact our technical specialists at m360@memmert.com.

Temperature

Working temperature range from 5°C (at least 20 below ambient temperature) to +90°C
Setting accuracy temperature up to 99.9 °C: 0.1 / from 100 °C: 0.5
Temperature sensor temperature measured through 4-wire Pt100 sensor individually for each thermoshelf
Setting temperature range from +5°C up to +90°C
resolution of display for actual values 0.1°C
resolution of display for setpoint values 0,1°C

Pressure (Vacuum)

Vacuum range 5 to 1100 mbar
Pressure control Digital electronic pressure control for a speed-controlled vacuum pump. Tubing for vacuum, air and inert gas are made of material 1.4571 (ASTM 316 Ti). Programmable, digitally controlled inlet for air.
Permitted final vacuum 0.01 mbar
Maximum leakage rate 0.01 bar/h
Pump control speed control, optimised rinsing procedures for the pump membranes as well as signal output for pump ON/OFF
Connection Vacuum connection with small flange DN16, and gas inlet with fresh air supply

Control technology

Calibrate Three freely selectable temperature values and three freely selectable pressure values
adjustable parameters temperature (Celsius or Fahrenheit), pressure (vacuum), programme time, time zones, summertime/wintertime
Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature control
Alarm visual and acoustic

Heating concept

Heating Concept Connection for a thermoshelf in the rear panel
VO cooling Peltier cooling unit ensures a surface temperature distribution with a maximum deviation of ±1 K across the entire temperature range

Standard equipment

Number of slide-in units 1
Works calibration certificate for +10°C/20 mbar
Door full-sight glass door, inside spring-loaded, 15 mm thick glazed panel in safety glass, outside with anti-splitter screen
Internals Cooling- and heating-combination of aluminium, material 3.3547 (ASTM B209), with integrated cooling circuit as well as large-area heating including local temperature sensing (Pt100, 4-wire-circuit); individual overtemperature protection. Further data see stainless steel interior.

Stainless steel interior

Material Hermetically welded stainless steel interior of extremely corrosion-resistant stainless steel, material 1.4404
Volume 49 l
Dimensions w(A) x h(B) x d(C): 385 x 385 x 330 mm
Max. number of internals 1
Max. loading of chamber 60 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Installation 4 adjustable appliance feet
Dimensions w(D) x h(E) x d(F): 550 x 687 x 480 mm (d +38mm door handle)

Electrical data

Voltage Electrical load 230 V, 50/60 Hz
approx. 600 W

Ambient conditions

Altitude of installation max. 2,000 m above sea level
Ambient temperature +5 °C to +40 °C
Set Up The distance between the wall and the rear of the chamber must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from the wall must not be less than 8 cm.
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 660 x 870 x 590 mm
Net weight approx. 95 kg
Gross weight carton approx. 116 kg

Standard units are safety-approved and bear the test marks

Let us help you find your Memmert

Send inquiry